Recall Z-1462-2008

Abbott Laboratories · initiated January 21, 2008

Terminated

Product

AxSYM Drugs of Abuse/Toxicology Assay Disk Version 8.0; List Number: 3D54-08; Abbott La boratories, Diagnostics Division, Abbott Park, IL 60064

Codes / lots: Lot Numbers: 59311P100 and 59312P100.

Reason for recall

Error Message: If the Positive and Negative Interpretation Cutoff parameters (116/117) are edited after installation of the revised assay file, VRTX error #0002 in Task 40 is generated and the instrument locks up when it attempts to report an AxSYM Amphetamine/Methamphetamine II patient result.

FDA-determined cause: Software change control

Action taken

Consignees were sent a Product Correction-Immediate Action Required letter on 1/21/08. The letter provided users with the steps to take to temporarily correct the problem until the recalling firm is able to send a corrected assay file. Users were requested to return a customer reply letter acknowledging receipt of the letter. For additional information, contact 1-877-4ABBOTT.

Details

Recall number
Z-1462-2008
Initiated
January 21, 2008
Posted by FDA
May 14, 2008
Terminated
December 1, 2009
Product code
DKZ – Enzyme Immunoassay, Amphetamine
Quantity
1,270 disks
Distribution
Worldwide Distribution including USA territory of Puerto Rico, and countries of Australia, Canada, Chile, Colombia, Germany, Mexico, New Zealand, South Korea, Taiwan, Thailand and Venezuela.
Recalling firm
Abbott Laboratories, Abbott Park, IL

Official FDA recall record