Recall Z-1472-2008

Nova Biomedical Corporation · initiated February 21, 2008

Terminated

Product

Varta Easypack XL Lithium Polymer Batteries (P/N 41772 and 42215) Used with: Nova Biomedical StatStrip Glucose Hospital Meter (P/N 42224) and Nova Biomedical StatSensor Creatinine Hospital Meter and EZ CHEM Creatinine Meter (P/N 9600). Statstrip Meter an In vitro diagnostic for use by health care professionals in the quantitative determination of glucose (GLU) in whole blood. EZ Chem In vitro diagnostic for the quantitative measurement of creatinine in capillary, venous or arterial blood.

Codes / lots: Batteries coded: XX07 and earlier The Nova Biomedical products that contain the lithium batteries include the following part numbers: Part # 41772 - Battery Assy, Part # 42215 - Battery Replacement pack, Part # 42224 - Kit Meter, Part # 43735 - Kit Meter EZ CHEM, Part # 43773 - Kit Meter, Part # 43811 - Battery Replacement Pack EZ CHEM, Part # 44013 - Kit Meter, Part # 44014 - Kit Meter and Part # 44251 - Kit Meter.

Reason for recall

Lithium batteries may fail if the glucose or creatinine meters have been dropped.

FDA-determined cause: Component design/selection

Action taken

Nova Biomedical notified consignees by telephone and followed by the Recall Notification letter for the Vata Easy Pack XL Lithium Polymer Battery, dated 2/2108 and mailed 2/26/08. Users are notified of the problem with the battery in the Nova Stat Strip Blood Glucose Hospital Meter System and is replacing batteries by on site visit. E-Z-EM , Inc. notified their customers by letter dated 3/17/08. Contact Nova Biomedical Corporation at 1-800-545-6682 for assistance.

Details

Recall number
Z-1472-2008
Initiated
February 21, 2008
Posted by FDA
September 23, 2008
Terminated
October 28, 2008
Product code
CGA – Glucose Oxidase, Glucose
Quantity
4432 batteries
Distribution
Worldwide Distribution --- including USA and countries of Canada, Taiwan, Hungary, Malaysia, Australia, Russia, Thailand, New Zealand, Austria, France, Germany, UK, China, South Africa, Croatian, Peru, and Czech Republic.
Recalling firm
Nova Biomedical Corporation, Waltham, MA

Official FDA recall record