Recall Z-1477-2025
Stryker Corporation · initiated February 25, 2025
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Hemopore 8cm nasal/sinus temporary wound dressing, Catalog Numbers 5400-020-208 and 5400-020-208ITL (OUS)
Codes / lots: 5400-020-208 UDI-DI: 07613327117233, Lot Numbers: 2022093020 2023011632 2023041321 2023062320 2023092220 2023112232 2024032120 2024052420 2022093021 2023011932 2023042120 2023062220 2023092932 2023112420 2024032820 2024052320 2022100720 2023012732 2023042121 2023073120 2023101220 2023112320 2024032920 2024060620 2022100620 2023013033 2023050320 2023073121 2023101221 2023120120 2024040520 2024060720 2022101320 2023020132 2023050220 2023081520 2023101921 2023113020 2024040420 2024062020 2022101420 2023020632 2023051620 2023081521 2023101920 2023120632 2024041820 2024062120 2022101720 2023022033 2023051720 2023082520 2023102721 2023121533 2024041920 2024070420 2022102120 2023030220 2023052620 2023082521 2023102720 2024010532 2024042632 2024070520 2022102720 2023030320 2023052520 2023083120 2023110320 2024020120 2024042520 2024071538 2022102820 2023032420 2023060120 2023090120 2023110220 2024020220 2024042620 2024071738 2022110220 2023032320 2023060220 2023090820 2023110733 2024030820 2024050821 2022110320 2023032920 2023060820 2023090620 2023110920 2024030720 2024050820 2022112932 2023033020 2023060920 2023091520 2023110921 2024031420 2024051520 2023010433 2023040732 2023061520…
Reason for recall
There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.
FDA-determined cause: Under Investigation by firm
Action taken
An Urgent Medical Device Recall notification letter dated 2/24/25 was sent to customers. Actions to be taken 1. Immediately review your inventory to locate and quarantine any affected products at your facility. 2. Return the enclosed Business Reply Form (BRF), even if the affected product is no longer in inventory, to the email Instruments.Recalls@stryker.com. 3. Upon receipt of the completed BRF, Stryker will provide a shipping label to return recalled product on-hand. 4. A replacement will be provided upon receipt of the recalled product. 5. Maintain awareness of this communication internally and inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients accordingly. Instruments.recalls@stryker.com. RESPONSE IS REQUIRED.
Details
- Recall number
- Z-1477-2025
- Initiated
- February 25, 2025
- Posted by FDA
- March 28, 2025
- Product code
- NHB – Polymer, Ear, Nose And Throat, Synthetic, Absorbable
- Quantity
- 182,344 total units
- Distribution
- US Nationwide distribution.
- Recalling firm
- Stryker Corporation, Portage, MI
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1469-2025 Class II | Otopore Cylinder Standard outer ear dressings, Catalog Numbers 5400-010-000 and…There is a potential for blister seals on the product to present a bubble on the seal area of the packaging.… | 02/25/2025Open, Classified |
| Z-1470-2025 Class II | Nasopore Standard 4cm fragmentable nasal dressing, Catalog Numbers 5400-010-004 and…There is a potential for blister seals on the product to present a bubble on the seal area of the packaging.… | 02/25/2025Open, Classified |
| Z-1471-2025 Class II | Nasopore Standard 8cm fragmentable nasal dressing, Catalog Numbers 5400-010-008 and…There is a potential for blister seals on the product to present a bubble on the seal area of the packaging.… | 02/25/2025Open, Classified |
| Z-1472-2025 Class II | Otopore Cylinder outer ear wound dressing, Firm Catalog Number 5400-020-000 and Forte…There is a potential for blister seals on the product to present a bubble on the seal area of the packaging.… | 02/25/2025Open, Classified |
| Z-1473-2025 Class II | Nasopore 8cm fragmentable nasal dressing, Firm Catalog Number 5400-020-008 and Forte…There is a potential for blister seals on the product to present a bubble on the seal area of the packaging.… | 02/25/2025Open, Classified |
| Z-1474-2025 Class II | Nasopore 4cm fragmentable nasal dressing, Firm Catalog Number 5400-020-004 and Forte…There is a potential for blister seals on the product to present a bubble on the seal area of the packaging.… | 02/25/2025Open, Classified |
| Z-1475-2025 Class II | Otopore Square outer ear wound dressing, Firm Catalog Number 5400-020-100 and Forte…There is a potential for blister seals on the product to present a bubble on the seal area of the packaging.… | 02/25/2025Open, Classified |
| Z-1476-2025 Class II | Nasopore FD fragmentable nasal dressing, Catalog Numbers 5400-020-108 and 5400-020-108ITLThere is a potential for blister seals on the product to present a bubble on the seal area of the packaging.… | 02/25/2025Open, Classified |
| Z-1478-2025 Class II | Nasopore Ex Firm 4cm fragmentable nasal dressing, Catalog Number 5400-030-004 and…There is a potential for blister seals on the product to present a bubble on the seal area of the packaging.… | 02/25/2025Open, Classified |
| Z-1479-2025 Class II | Nasopore Ex Firm 8cm fragmentable nasal dressing, Catalog Number 5400-030-008 and…There is a potential for blister seals on the product to present a bubble on the seal area of the packaging.… | 02/25/2025Open, Classified |
| Z-1480-2025 Class II | Nasopore 4cm Standard 2PK fragmentable nasal dressing, Catalog Number 5400-212-004S and…There is a potential for blister seals on the product to present a bubble on the seal area of the packaging.… | 02/25/2025Open, Classified |
| Z-1481-2025 Class II | Nasopore 4cm Standard 2PK fragmentable nasal dressing, Catalog Number 5400-212-004S and…There is a potential for blister seals on the product to present a bubble on the seal area of the packaging.… | 02/25/2025Open, Classified |
| Z-1482-2025 Class II | Hemopore 2PK nasal/sinus temporary wound dressing, Catalog Number 5400-222-208SThere is a potential for blister seals on the product to present a bubble on the seal area of the packaging.… | 02/25/2025Open, Classified |