Recall Z-1478-2014

BioMerieux SA · initiated March 6, 2014

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

API Listeria (REF 10300). ZYM B reagent is used and included inside of API NH (REF 10400) and API Listeria (10300). All of the APIs are standardized systems combining several biochemical tests that enable group or species identification of most microorganisms.

Codes / lots: Code - 1002569040, Exp. Date 8/29/2014, 1002569041, Exp. Date 8/296/2014, 1002587190, Exp. Date 9/3/2014, 1002856790, Exp. Date 7/18/2014, 1002856800, Exp. Date 6/12/2014

Reason for recall

bioMerieux identified a visual defect and activity issue on the ZYM B reagent (REF 70493) and FB reagent (REF 70562). The activity issue lead to a false negative result of some Rapid ID 32 strips and API biochemical tests using the ZYM B reagent or FB reagent

FDA-determined cause: Other

Action taken

Letters were issued to customers on March 10, 2014. The letter contained instructions to be used if they still were using the product. The letter also included a report form which was to be returned.

Details

Recall number
Z-1478-2014
Initiated
March 6, 2014
Posted by FDA
April 18, 2014
Terminated
March 14, 2017
Product code
LIB – Device, General Purpose, Microbiology, Diagnostic
Quantity
33,853
Distribution
Worldwide Distribution-USA (nationwide) including DC and Puerto Rico, and the states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WV, WA, WI, and WY and the country of Algeria, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Benin…
Recalling firm
BioMerieux SA, Marcy L'Etoile

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1475-2014 Class IIZYM B Reagent (REF 70493). The ZYM B reagent is an additional test used to reveal the…bioMerieux identified a visual defect and activity issue on the ZYM B reagent (REF 70493) and FB reagent… 03/06/2014Terminated
Z-1476-2014 Class IIFB Reagent (REF 70562). FB reagent is an additional test used for revealing the results…bioMerieux identified a visual defect and activity issue on the ZYM B reagent (REF 70493) and FB reagent… 03/06/2014Terminated
Z-1477-2014 Class IIAPI NIH (REF 10400). ZYM B reagent is used and included inside of API NH (REF 10400) and…bioMerieux identified a visual defect and activity issue on the ZYM B reagent (REF 70493) and FB reagent… 03/06/2014Terminated