Recall Z-1483-2021

Welch Allyn Inc Mortara · initiated March 31, 2021

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

ELI 380 Electrocardiograph, with AM12M Acquisition Module accessory. Impacted software version of the AM12M is version 1.5.1 - Product Usage: indicated for use on adult and pediatric populations.

Codes / lots: Part Number: ELI380-ACX32 (Device Identifier: 732094265415); S/N: 118490001284, 119150000588, 120090001355, 120090001354 Part Number: ELI380-DAX3X (Device Identifier: 812345026757); S/N: 117120557018, 117120557019, 117120557020, 117120557021, 117230571942, 117230571943, 117270577469 Part Number: ELI380-DBX32 (Device Identifier: 732094265262); S/N: 118180634420 Part Number: ELI380-DCX32 (Device Identifier: 732094265187); S/N: 118340001074, 118340001074, 118340001075, 118500001332, 118460001178, 119150000120, 119210000774, 119290000580, 119520000194, 120080001509, 120080001510, 120100000310, 120080001511, 120110002478, 120170000570

Reason for recall

Devices do not meet IEC 60601-2-27 requirements as labeled.

FDA-determined cause: Process control

Action taken

Recall notification letters were mailed to customers notifying them of the potential issue with recommended actions beginning April 7, 2021. The notification letter includes an acknowledgement form that must be returned to the recalling firm. Once the customer confirms the impacted units are in possession, the recalling firm will arrange for an exchange AM12M to be provided. The recalled unit will be returned to the recalling firm's service center for processing.

Details

Recall number
Z-1483-2021
Initiated
March 31, 2021
Terminated
August 16, 2024
Product code
DPS – Electrocardiograph
Quantity
27 units
Distribution
US Nationwide distribution in the states of CA, CO, FL, GA, IN, LA, NC, NJ, NV, NY, OH, OK, TX, UT, and VA.
Recalling firm
Welch Allyn Inc Mortara, Milwaukee, WI

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1482-2021 Class IIELI 280 Electrocardiograph, with AM12M Acquisition Module accessory. Impacted software…Devices do not meet IEC 60601-2-27 requirements as labeled. 03/31/2021Terminated
Z-1484-2021 Class IISurveyor S12/S19 Bedside Patient Monitor, with AM12M Acquisition Module accessory.…Devices do not meet IEC 60601-2-27 requirements as labeled. 03/31/2021Terminated
Z-1485-2021 Class IIAM12M Acquisition Module accessory. Impacted software version of the AM12M is version…Devices do not meet IEC 60601-2-27 requirements as labeled. 03/31/2021Terminated