Recall Z-1485-2015

Wright Medical Technology, Inc. · initiated March 6, 2015

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

CROSSCHECK(R) Plating system, LAPIDUS CX Plate, NEUTRAL, Right, REF CCP-LPX0R. Orthopedic use.

Codes / lots: Lot Numbers: 35049-31A-1, 35049-3A, 35049-3A-1-1, 362812, 36281-2-1

Reason for recall

The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunction

FDA-determined cause: Nonconforming Material/Component

Action taken

Firm has initiated the recall on 03/06/2015 through written communication via Fed Ex P1 delivery to all Wright Medical Technologies distributors (hubs). Follow-up to non-responding consignees will commence 30 days after initiation. Effectiveness checks will be performed through the confirmation of returned product via a validated ERP system, a returned/signed verification form, and in the case of non-responders, a phone call, e-mail or follow up letter, depending on the region and availability of consignee.

Details

Recall number
Z-1485-2015
Initiated
March 6, 2015
Posted by FDA
April 22, 2015
Terminated
August 5, 2018
Product code
HRS – Plate, Fixation, Bone
Quantity
102 units
Distribution
Nationwide Distribution.
Recalling firm
Wright Medical Technology, Inc., Memphis, TN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1482-2015 Class IICROSSCHECK(R) Plating system, HOCKEY STICK CX Plate, Left, REF CCP-HSX1L. Orthopedic use.The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification… 03/06/2015Terminated
Z-1483-2015 Class IICROSSCHECK(R) Plating system, HOCKEY STICK CX Plate, Right, REF CCP-HSX1R. Orthopedic use.The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification… 03/06/2015Terminated
Z-1484-2015 Class IICROSSCHECK(R) Plating system, LAPIDUS CX Plate, NEUTRAL, Left, REF CCP-LPX0L. Orthopedic…The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification… 03/06/2015Terminated
Z-1486-2015 Class IICROSSCHECK(R) Plating system, LAPIDUS CX Plate, 1mm Step, Left, REF CCP-LPX1L.…The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification… 03/06/2015Terminated
Z-1487-2015 Class IICROSSCHECK(R) Plating system, LAPIDUS CX Plate, 1mm Step, Right, REF CCP-LPX1R.…The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification… 03/06/2015Terminated
Z-1488-2015 Class IICROSSCHECK(R) Plating system, LAPIDUS CX Plate, 2mm Step, Left, REF CCP-LPX2L.…The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification… 03/06/2015Terminated
Z-1489-2015 Class IICROSSCHECK(R) Plating system, LAPIDUS CX plate, 2mm Step, Right, REF CCP-LPX2R.…The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification… 03/06/2015Terminated
Z-1490-2015 Class IICROSSCHECK(R) Plating system, MTP NX Plate, Left, Recon, REF CCP-MPN2L. Orthopedic use.The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification… 03/06/2015Terminated
Z-1491-2015 Class IICROSSCHECK(R) Plating system, MTP NX Plate, Right, Recon, REF CCP-MPN2R. Orthopedic use.The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification… 03/06/2015Terminated
Z-1492-2015 Class IICROSSCHECK(R) Plating system, MTP CX Plate, Left, REF CCP-MPX1L. Orthopedic use.The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification… 03/06/2015Terminated
Z-1493-2015 Class IICROSSCHECK(R) Plating system, MTP CX Plate, Right, REF CCP-MPX1R. Orthopedic use.The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification… 03/06/2015Terminated
Z-1494-2015 Class IICROSSCHECK(R) Plating system, Utility Plate, 2-Hole, REF CCP-UTN2. Orthopedic use.The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification… 03/06/2015Terminated
Z-1495-2015 Class IICROSSCHECK(R) Plating system, Utility Plate, 3-Hole, REF CCP-UTN3. Orthopedic use.The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification… 03/06/2015Terminated
Z-1496-2015 Class IICROSSCHECK(R) Plating system, Utility Plate, 4-Hole, REF CCP-UTN4. Orthopedic use.The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification… 03/06/2015Terminated
Z-1497-2015 Class IICROSSCHECK(R) Plating system, Utility Plate, 5-Hole, REF CCP-UTN5. Orthopedic use.The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification… 03/06/2015Terminated
Z-1498-2015 Class IICROSSCHECK(R) Plating system, Utility Plate, 6-Hole, REF CCP-UTN6. Orthopedic use.The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification… 03/06/2015Terminated
Z-1499-2015 Class IICROSSCHECK(R) Plating system, Utility Plate, 7-Hole, REF CCP-UTN7. Orthopedic use.The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification… 03/06/2015Terminated
Z-1500-2015 Class IICROSSCHECK(R) Plating system, Y-PLATE CX, REF CCP-YPX1. Orthopedic use.The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification… 03/06/2015Terminated