Recall Z-1494-2021

Coltene Whaledent Inc · initiated March 18, 2021

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Dental pins - TMS LINK PLUS Self-Threading Pins - Product Usage: implanted into the teeth or for lab use for dental restoration purposes. The normal procedure requires a hole to be drilled into the tissue of a tooth with an intraoral drill prior to installation. Model number/Catalog number: L741

Codes / lots: Lot/Batch number: K39841

Reason for recall

The incorrect dental pins were packaged and subsequently distributed.

FDA-determined cause: Under Investigation by firm

Action taken

On 3/18/2021, Coltene issued an Urgent Medical Device Recall notice via letter to customers notifying them the incorrect dental pins were packaged and subsequently distributed. The TMS LINK PLUS Self-Threading L741 Pins should be stainless steel, single sheer link pins that are gold-plated. The pins that were packed and distributed are titanium, double sheer, and not gold plated. The incorrectly packaged pin is too small for the drill, resulting in a hole that may be drilled too large for the pins. The pins may not engage the hole to effectively sheer.

Details

Recall number
Z-1494-2021
Initiated
March 18, 2021
Product code
EBL – Pin, Retentive And Splinting, And Accessory Instruments
Quantity
65 packs of 20 pins
Distribution
Worldwide distribution - US Nationwide distribution in the states of NJ, IN, NC, IL. GA, CA, NY, PA and the countries of Canada, Venezuela.
Recalling firm
Coltene Whaledent Inc, Cuyahoga Falls, OH

Official FDA recall record