Recall Z-1498-2016

Optovue, Inc. · initiated March 16, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

iVue and iVue 500 (now called iScan) OCT systems with software version 2016.0.0.127 (USA only) and 2016.1.0.127 (out of US) Opthalmic: iVue is a non-contact, high resolution tomographic imaging device. It is intended for in vivo imaging, axial cross-sectional and three-dimensional imaging and measurement of anterior and posterior ocular structures, including retina, retinal nerve fiber layer, ganglion cell complex (GCC), optic disc, cornea, and anterior chamber the eye.

Codes / lots: US and foreign (not yet provided) Software version 2016.0.0.127: Serial numbers: 21908, 26507, 26496, 26558, 22738, 26139, 26442, 25720, 25079, 25142, 23298, 21990, 23461, 24199, 25843, 26572, 26550, 22971, 26495, 22652, 26550, 22971, 26495, 22652, 25389, 23309, 23760, 21732-1, 23527, 20923, 21600, 21414, 20271, 23153, 22875, 23211. Software version 2016.1.0.127; Serial numbers forthcoming.

Reason for recall

A software anomaly was detected that affects Optovue iVue and iVue 500 with Normative Database that may display images of the same eye rather than right and left if user scrolls on the map.

FDA-determined cause: Under Investigation by firm

Action taken

Recall Letters were sent to customers on March 22, 2016 to inform them of the recall.

Details

Recall number
Z-1498-2016
Initiated
March 16, 2016
Terminated
April 4, 2017
Product code
HLI – Ophthalmoscope, Ac-Powered
Quantity
40
Distribution
PA, CA, IL, HI. MO, CO, GA, ME, IN, AZ, WV, VA, FL, AR, UT, NE, IN, NV, LA, TN, KS, OK. Foreign distribution forthcoming
Recalling firm
Optovue, Inc., Fremont, CA