Recall Z-1509-2017

Ohmeda Medical · initiated November 18, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Giraffe OmniBed Ohmeda Medical Omnibed- The OmniBed is a combination of an infant incubator ( incubator ) and an infant radiant warmer ( warmer ). Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperature controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate s transition from the uterus to the external environment

Codes / lots: See Attached Spreadsheat

Reason for recall

Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.

FDA-determined cause: Under Investigation by firm

Action taken

According to the firm, this issue will be resolved by sending an Urgent Medical Device Correction customer letter to customers with affected Power Cords. This letter will inform the customer of the issue and provide safety instructions for immediate mitigation of the issue. The long-term solution is to re-place all affected Power Cords at no cost to the customer. According to the firm, the plan for public notification is to contact all customers with affected Power Cords by means of an Urgent Medical Device Correction letter. In the US this will be sent by a traceable means (either US Postal Service with receipt acknowledgement (registered mail) or equivalent courier service such as FEDEX). This letter will be sent to the following titles within the affected accounts; Nurse Managers, Labor & Delivery/NICU, Bio-Medical Engineering Department Managers and Risk Management Directors We will track all service requests for the field action to completion. These service requests will record the replacement of the affected Power Cords. Exceptions to completion of the service request will be recorded as well. Records of field action completion will be retained per our processes and procedures.

Details

Recall number
Z-1509-2017
Initiated
November 18, 2016
Terminated
July 13, 2018
Product code
FMZ – Incubator, Neonatal
Quantity
5,993 (5,080 US; 913 OUS)
Distribution
Worldwide Distribution - US (nationwide and Puerto Rico) Bahamas and Internationally to UK, Trinidad & Tobago, Taiwan, Spain, Panama, Netherlands Antilles, Mexico, Lebanon, Japan, Ireland, Honduras, El Salvador, Ecuador, Costa Rica, Colombia, Canada, Brazil, and Venezuela,
Recalling firm
Ohmeda Medical, Laurel, MD

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1501-2017 Class IIGiraffe Exam Light The Exam Light illuminates body surfaces and facilitates patient…Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are… 11/18/2016Terminated
Z-1502-2017 Class IIGiraffe Incubator The Giraffe Incubator is an infant incubator. Incubators provide heat…Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are… 11/18/2016Terminated
Z-1503-2017 Class IIGiraffe Spot PT Ohmeda Medical Spot PT Lite Phototherapy System- The Spot PT Lite…Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are… 11/18/2016Terminated
Z-1504-2017 Class IIGiraffe OmniBed Ohmeda Medical Giraffe OmniBed- The OmniBed is a combination of an…Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are… 11/18/2016Terminated
Z-1505-2017 Class IIGiraffe Incubator Ohmeda Medical Giraffe Incubator-The Giraffe Incubator is an infant…Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are… 11/18/2016Terminated
Z-1506-2017 Class IIGiraffe Bedded Warmer, Panda Warmer Giraffe and Panda Warmer- Infant radiant warmers…Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are… 11/18/2016Terminated
Z-1507-2017 Class IIGiraffe Bedded Warmer, Panda Warmer Modification to Giraffe and Panda Warmers (Giraffe…Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are… 11/18/2016Terminated
Z-1508-2017 Class IIGiraffe Bedded Warmer, Panda Warmer Modifiction to Giraffe and Panda Warmer- Infant…Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are… 11/18/2016Terminated
Z-1510-2017 Class IICarePlus Incubator Ohmeda-Ohio CarePlus Incubator- Incubators provide a controlled…Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are… 11/18/2016Terminated
Z-1511-2017 Class IIIWS Infant Warmer System-Infant radiant warmers provide infrared heat, in a controlled…Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are… 11/18/2016Terminated