Recall Z-1509-2020
Stryker Corporation · initiated September 12, 2019
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Stryker, REF: 0250070450, 5.0 mm x 45 cm, Strykeprobe, Outer Sheath Replacement, RX Only, Non-Sterile, CE, UDI: (01) 07613327051926 This products was distributed with the following sheath/probe: 0250070451-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, Spatula-Tip, 0250070452-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, J-TIP, 0250070453-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, L-TIP, 0250070455-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, Needle Tip
Codes / lots: Lot Numbers: 010818-02, 010818-05, 010818-06, 010818-07, 010818-08, 010818-09, 040816-01, 040816-02
Reason for recall
Potential diff. in the length of the sheath tube and the sheath/probe. A diff. in length could lead to the base of the sheath/probe not being fully seated in the sheath tube which could result in the distal tip of the sheath melting. This could lead to a compromised insulation, leading to unintentional flow of electricity to the patient. The sheath/probe is packaged separately.
FDA-determined cause: Other
Action taken
On September 25, 2019 and November 26, 2019, the firm sent an "Urgent Medical Device Recall Notification" letter to all consignees via USPS mail. In addition, to informing consignees about the recall, they recall letter asked consignees to take the following actions: Actions to be taken by the Customer/User: 1. Inform individuals within your organization who need to be aware of this device removal. 2. Check all stock areas and/or operating room storage to determine if any devices with the affected Strykeprobe Sheath lot numbers from Attachment A are at your facility. - a. If you would like a list of all affected probe lot numbers, please email EndoRecall@Stryker.com 3. If affected product is found, segregate the product and call Stryker customer service at 1-800-624-4422 (Option 3) or email endocustomersupport@stryker.com to arrange for product return and issuance of credit or replacement (upon availability). - a. Remove the affected sheath, quarantine, and discontinue use of the identified non-conforming recalled sheath(s). - b. When returning units, please enclose Business Reply Form on Attachment B. 4. If affected product is NOT found: - a. Please complete Business Reply Form on Attachment B and return to EndoRecall@Stryker.com 5. Please forward a copy of this letter to any other personnel within your facility you deem appropriate. 6. Please send any questions to EndoRecall@Stryker.com.
Details
- Recall number
- Z-1509-2020
- Initiated
- September 12, 2019
- Terminated
- October 31, 2022
- Product code
- GEI – Electrosurgical, Cutting & Coagulation & Accessories
- Distribution
- OUS: Netherlands, Spain, Switzerland, Colombia, Canada, Japan, Taiwan,, Puerto Rico, South Africa, and South Korea
- Recalling firm
- Stryker Corporation, San Jose, CA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1508-2020 Class II | Stryker, REF: 0250070460, 5.0 mm x 32 cm, Strykeprobe, Outer Sheath Replacement, RX…Potential diff. in the length of the sheath tube and the sheath/probe. A diff. in length could lead to the… | 09/12/2019Terminated |