Recall Z-1519-2012
Instrumentation Laboratory Co. · initiated April 11, 2012
Product
Instrumentation Laboratory (IL)HemosIL von Willebrand Factor Activity Assay Kit Part Number: 0020004700 The assay is intended for the quantitative determination of von Willebrand Factor Activity (VWF Activity) in human citrated plasma on IL Coagulation Systems
Codes / lots: Lot Numbers: B01424 Exp Date: March 31 , 2012; B11498 Exp Date: July, 31 , 2012; B11655 Exp Date: November 30, 2012; B11687 Exp Date: January 31 , 2013
Reason for recall
Falsely elevated levels may be reported with the HemosIL von Willebrand Factor Activity assay.
FDA-determined cause: Nonconforming Material/Component
Action taken
Instrumentation Laboratories mailed Recall Letter via USPS to US and Canadian customers on April 11, 2012, through Beckman Coulter. The customer mailing and tracking of the Urgent Recall Notification and Recovery Procedure in the United States and Canada will be handled through the distributor, Beckman Coulter. Accounts are requested to check inventory for the product lots of Hemosll von Willebrand Factor Activity, PN 0020004700, listed in the table and document their destruction on the provided tracking form and return the completed form. Questions or concerns regarding this notification, please call Beckman Coulter Technical Service at (800) 526-7694. On March 29, an e-mail communication was circulated to the IL Affiliates in Europe and Japan requesting them to translate the Urgent Field Safety Notification [title of document outside US and Canada] and product destruction tracking form, mail the notification to their customer base and notify their Competent Authorities.. Fpr questions regarding this recall call 781-861-4467.
Details
- Recall number
- Z-1519-2012
- Initiated
- April 11, 2012
- Posted by FDA
- May 10, 2012
- Terminated
- August 24, 2012
- Product code
- GGP – Test, Qualitative And Quantitative Factor Deficiency
- Quantity
- 9320 Kits
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Canada, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELARUS BELGIUM, BRAZIL, CHILE, CHINA, CZECH REPUBLIC, DENMARK FRANCE, GERMANY, GREECE, HUNGARY, INDIA, IRAN (ISLAMIC REPUBLIC OF), ISRAEL, ITALY, JAPAN, JORDAN, KOREA, REPUBLIC OF KUWAIT LITHUANIA, LUXEMBOURG, MALAYSIA, MEXICO, MOROCCO, NETHERLANDS NORWAY, PERU, PHILIPPINES, POLAND…
- Recalling firm
- Instrumentation Laboratory Co., Bedford, MA