Recall Z-1519-2020
Avid Medical, Inc. · initiated February 18, 2020
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Custom procedure trays containing medical components needed to facilitate patient procedures and to perform minor and major patient surgeries. Multiple item and lot numbers.
Codes / lots: Item Codes: SHER010-12. Lot Numbers: 1339900, 1352089, 1361835, 1370391, 1372029. Expiration dates: 12/15/2020-09/05/2021. ,
Reason for recall
The medical procedure trays contain one or more recalled surgical gowns. The gowns were manufactured at locations that did not maintain proper environmental conditions. There is no assurance that the surgical gowns are sterile. An inadequately sterilized surgical gown could compromise a sterile field and increase the risk of a surgical site infection. Due to the configuration of the trays, the recalled gowns may come into contact with other components within the tray posing a risk of cross contamination.
FDA-determined cause: Other
Action taken
URGENT MEDICAL DEVICE RECALL NOTICE letters, dated February 18, 2020, were distributed to customers. Letter provides identification of the issue, health risk and action to take. Customers were instructed to take the following required actions: " Review all storage and usage locations and segregate and quarantine any trays with an item code and lot number listed on attached Exhibit A. " Complete and return the enclosed Recall Response Form to Sharon.moll@owens-minor .com. Please respond even if you do not have any trays subject to this recall. " Upon receipt of your completed Recall Response Form, product return instructions will be provided and a RGA will be issued, if you do have trays subject to this recall. " If you have further distributed this product, please identify your customers and notify them at once of this product recall. Your notification to your customers may be enhanced by including a copy of this Urgent Medical Device Recall Notice . Your assistance is appreciated and necessary to prevent any unintended use of the product. If you have any questions, call Sharon Moll at 757-566-9844.
Details
- Recall number
- Z-1519-2020
- Initiated
- February 18, 2020
- Terminated
- November 23, 2020
- Product code
- OGJ – Nerve Block Tray
- Quantity
- 230 each
- Distribution
- US: AZ, CA, IL, IN, MD, MS, NJ, NV, NY, TN, TX WI, VA
- Recalling firm
- Avid Medical, Inc., Toano, VA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1515-2020 Class II | Custom procedure trays containing medical components needed to facilitate patient…The medical procedure trays contain one or more recalled surgical gowns. The gowns were manufactured at… | 02/18/2020Terminated |
| Z-1516-2020 Class II | Custom procedure trays containing medical components needed to facilitate patient…The medical procedure trays contain one or more recalled surgical gowns. The gowns were manufactured at… | 02/18/2020Terminated |
| Z-1517-2020 Class II | Custom procedure trays containing medical components needed to facilitate patient…The medical procedure trays contain one or more recalled surgical gowns. The gowns were manufactured at… | 02/18/2020Terminated |
| Z-1518-2020 Class II | Custom procedure trays containing medical components needed to facilitate patient…The medical procedure trays contain one or more recalled surgical gowns. The gowns were manufactured at… | 02/18/2020Terminated |
| Z-1520-2020 Class II | Custom procedure trays containing medical components needed to facilitate patient…The medical procedure trays contain one or more recalled surgical gowns. The gowns were manufactured at… | 02/18/2020Terminated |
| Z-1521-2020 Class II | Custom procedure trays containing medical components needed to facilitate patient…The medical procedure trays contain one or more recalled surgical gowns. The gowns were manufactured at… | 02/18/2020Terminated |
| Z-1522-2020 Class II | Custom procedure trays containing medical components needed to facilitate patient…The medical procedure trays contain one or more recalled surgical gowns. The gowns were manufactured at… | 02/18/2020Terminated |
| Z-1523-2020 Class II | Custom procedure trays containing medical components needed to facilitate patient…The medical procedure trays contain one or more recalled surgical gowns. The gowns were manufactured at… | 02/18/2020Terminated |