Recall Z-1522-2008

Datex - Ohmeda, Inc · initiated May 19, 2008

Terminated

Product

GE Healthcare Innova 4100 Digital Fluoroscopic Imaging (Cardiovascular Imaging System

Codes / lots: Systems id: 77573841, 80158241, 910063039, 3018774100, 4103374100, 4438494100, 5134754100, 5138534100, 6029234100, 8159374100, 8595724100, 9408984100, 82406010011, 082416130011, 82416130012, 00095VAS07, 205759CL2, 216844INNOVA, 229228INNOV41V, 229434INNOV, 239436CATH1, 251435ICV5, 304255INNOVA, 304347INNOVA, 315448RM4, 315448RM5, 319338VA4100, 325670CATH, 337981INNOVA, 352333VASC, 352376SP, 352401CATH, 352401VASC, 407518SP2, 407518SP3, 407SPECIALS, 414961INN4100, 478633INOV, 508856OR41, 515263IL4100, 516562INNOVA4, 573632CRA, 573761SMA, 574535IN41, 600050VA01, 6012884100SP1, 606833INNOVA, 609652CATH3, 614293LIBERTY1, 614293LIBERTY2, 704487CR41, 706802FINN, 715369INN4100, 717763INNOVA41, 717763OR4100, 718518CVL1, 724656ANG, 727725SPEC1, 727819SPSC, 740592INNOVA, 757398INNOVA, 757736INNOVA, 803329CL41, 804320CATH2B, 804594CATH, 812279INNOVA1, 813972INTV, 816276SP1, 817255NH4100, 828696IN41, 865541CCL5, 865693CCL3, 901726SP2, 903CSM4100, 904276SP, 906485ANGIO, 918307SHIN4, 941483INNOVA41, 956632INN4100, A5630803, B5801711, BE5005VA01, BY4020VA01, RU1018VA03, SA1188VA03, SLIN4100, YV0003.

Reason for recall

GE Healthcare has become aware of a potential condition that may impact operator or patient safety. One customer has reported that inability to terminate X-ray exposure after releasing the hand switch control (located in the control room).

FDA-determined cause: Software design

Action taken

An Urgent Medical Device Correction letter, dated June 2008 was sent to customers The letter described the issue, affected products and provided short term user recommendations. GE developed a new software release that corrects this condition and will launch the corrective action to upgrade all affected systems.

Details

Recall number
Z-1522-2008
Initiated
May 19, 2008
Posted by FDA
September 30, 2008
Terminated
June 30, 2016
Product code
MQB – Solid State X-Ray Imager (Flat Panel/Digital Imager)
Quantity
87
Distribution
Worldwide Distribution - AL, AZ, CA, FL,, GA, HI, IA,IL, IN, KY, LA,MA, MD, MI, MO, MS, NC, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WI, and WV, Puerto Rico. OUS to include: Australia, Byelarus, China, Denmark, France, Germany, Greece, Honduras, Iran, Iraq, Italy, Japan, Latvia, Lebanon, Mexico, Poland, Russia, Saudi Arabia, Turkey, United Kingdom, and Venezuela.
Recalling firm
Datex - Ohmeda, Inc, Madison, WI

Official FDA recall record