Recall Z-1524-2013

Bard Peripheral Vascular Inc · initiated April 16, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

BARD Biopsy Systems UltraClip Dual Trigger breast tissue marker. Intended for use to attach to soft breast tissue at the surgical site during an open surgical breast biopsy.

Codes / lots: Affected Product Lots with the following Product Code and Lot Number combinations. Product Code 864017D 864017D 864017D 864017D 864017D 864017D 864017D 864017DL Lot Number HUWI1522 HUWI1941 HUWI1942 HUWI1943 HUWJ1851 HUWJ1852 HUWK0350 HUWJ1610

Reason for recall

Bard Peripheral Vascular (BPV) has confirmed that some devices with the recalled product code / lot number combination may have the metal tissue marker separated from the needle prior to use.

FDA-determined cause: Nonconforming Material/Component

Action taken

Bard Peripheral Vascular, Inc. initiated a recall of certain product code and lot number combinations of Bard UltraClip Dual Trigger Tissue Marker by contacting customers on 04/16/2013. On 04/18/2013, BPV sent a Customer Recall notification Letter via FedEx with proof of delivery notice. The letter included reason for recall, a table of affected product code and lot numbers, instructions (Do not use or further distribute any affected product), a response form (titled Recall and Effectiveness Check Form) enclosed, and contact information.

Details

Recall number
Z-1524-2013
Initiated
April 16, 2013
Posted by FDA
June 8, 2013
Terminated
January 30, 2014
Product code
FZP – Clip, Implantable
Quantity
3,085
Distribution
Worldwide Distribution: USA Nationwide and Internationally to Europe.
Recalling firm
Bard Peripheral Vascular Inc, Tempe, AZ

Official FDA recall record