Recall Z-1530-2024

Cook Incorporated · initiated March 1, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Dilator - Intended to be used for dilating puncture sites or catheter tracts for percutaneous placement of devices for vascular and non-vascular applications such as in the venous, arterial, biliary and renal systems. 1) JCD10.0-38-20 2) JCD20.0-38-20 3) JCD22.0-38-20 4) JCD6.0-38-15 5) JCD8.0-38-15 6) JCD8.0-38-20

Codes / lots: 1)JCD10.0-38-20 (01)00827002009936(17) Lot Numbers: 15794371, 15794373, 15794374, 15806068, 15816198, 15816199 15816200 2) JCD20.0-38-20 (01)00827002014718(17) Lot Numbers: 15784153, 15784159, 15784160, 15794192, 15794199, 15794202. 15804028 3) JCD22.0-38-20 (01)00827002014725(17) Lot Numbers: 15794207, 15806062 4) JCD6.0-38-15 (01)00827002009448(17) Lot Number: 15814407 5) JCD8.0-38-15 (01)00827002009998(17) Lot Numbers: 15797971, 15797979 6) JCD8.0-38-20 (01)00827002009806(17) Lot Number:15791073

Reason for recall

Packaging may have low seal strength and not meet peel strength specifications, compromising device sterility

FDA-determined cause: Under Investigation by firm

Action taken

Cook Medical issued Urgent Medical Device Recall letter on 3/1/24 to Risk Management/Recall Administration. . Letter states reason for recall, health risk and action to take: Actions to be Taken by the Customer 1. Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all further distribution and use of the affected products. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter regardless of whether you have affected product(s) on hand. You may return the Form via fax (812.339.7316) or email (FieldActionsNA@CookMedical.com). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Please immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com. Action being Taken by Cook Medical Cook Medical is removing potentially affected products from the market. Additionally, a supplier investigation has been opened to determine the appropriate corrective action(s) to prevent reoccurrence of a similar issue.

Details

Recall number
Z-1530-2024
Initiated
March 1, 2024
Posted by FDA
April 12, 2024
Product code
FGE – Stents, Drains And Dilators For The Biliary Ducts
Quantity
5,998 units
Distribution
Worldwide distribution - US Nationwide and the countries of AT, AU, BR, CA, CH, DE, ES, GB, HK, IE, IL, IN, IT, NZ, SG.
Recalling firm
Cook Incorporated, Bloomington, IN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1525-2024 Class IIChiba Biopsy Needle - Used for aspiration biopsy. REFERENCE PART NUMBER (RPN) ORDER…Packaging may have low seal strength and not meet peel strength specifications, compromising device sterility 03/01/2024Open, Classified
Z-1526-2024 Class IITrocar Needle - Used for dilating puncture sites or catheter tracts for percutaneous…Packaging may have low seal strength and not meet peel strength specifications, compromising device sterility 03/01/2024Open, Classified
Z-1527-2024 Class IIYueh Centesis Disposable Catheter Needle - Intended for use in percutaneous fluid…Packaging may have low seal strength and not meet peel strength specifications, compromising device sterility 03/01/2024Open, Classified
Z-1528-2024 Class IIPeel-Away Introducer - Intended for the percutaneous introduction of balloon, electrode…Packaging may have low seal strength and not meet peel strength specifications, compromising device sterility 03/01/2024Open, Classified
Z-1529-2024 Class IIEntuit Secure Adjustable Gastrointestinal Suture Anchor Set - Intended for anchoring the…Packaging may have low seal strength and not meet peel strength specifications, compromising device sterility 03/01/2024Open, Classified
Z-1531-2024 Class IIDilator With AQ Hydrophilic Coating - Intended to be used for dilating puncture sites or…Packaging may have low seal strength and not meet peel strength specifications, compromising device sterility 03/01/2024Open, Classified
Z-1532-2024 Class IIDilator - Coons Taper - Used for dilating puncture sites or catheter tracts for…Packaging may have low seal strength and not meet peel strength specifications, compromising device sterility 03/01/2024Open, Classified