Recall Z-1534-2013

OrthoPediatrics Corp · initiated April 19, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

90 Degree Cannulated Infant Blade Plate 25mm x 5mm x 3 Hole Product Usage: The OrthoPediatrics Locking Cannulated Blade Plate System is intended for fixation of long bone fractures and osteotomies in all pediatric subgroups (except neonates) and in small stature adults. Specific indications include: intertrochanteric derotation and varus osteotomies, femoral neck and pertrochanteric fractures, intertrochanteric valgus osteotomies, proximal and distal tibial osteotomies and humeral fractures and osteotomies.

Codes / lots: All lot numbers are affected. Catalog No. 00-1200-1000

Reason for recall

Post Market Surveillance of the 90 degree Cannulated Infant Blade Plate indicated that there have been five reports concerning bending of the plate intra-operatively.

FDA-determined cause: Device Design

Action taken

OrthoPediatrics sent an Urgent Voluntary Recall notifiation via telephone and email to all affected consignees. The letter instructed consignees to take possession of any affected devices and quarantine immediately. A Mandatory Reply Form is included with the notification letter to be completed and returned. Please contact Greg Teghtmeyer if you have any questions (574-268-6379).

Details

Recall number
Z-1534-2013
Initiated
April 19, 2013
Posted by FDA
June 12, 2013
Terminated
September 19, 2013
Product code
HWC – Screw, Fixation, Bone
Quantity
130 devices were shipped
Distribution
USA Nationwide Distribution
Recalling firm
OrthoPediatrics Corp, Warsaw, IN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1535-2013 Class II90 Degree Cannulated Infant Blade Plate 30mm x 5mm x 3 Hole Product Usage: The…Post Market Surveillance of the 90 degree Cannulated Infant Blade Plate indicated that there have been five… 04/19/2013Terminated
Z-1536-2013 Class II90 Degree Cannulated Infant Blade Plate 35mm x 5mm x 3 Hole Product Usage: The…Post Market Surveillance of the 90 degree Cannulated Infant Blade Plate indicated that there have been five… 04/19/2013Terminated