Recall Z-1537-2012
Haemacure Corporation · initiated November 16, 2009
Product
HemaSyst, Dual Cannula Tip with Adapter 16 ga. x 4" (10.2 cm). Product is labeled in part:"***Pouch label***HAEMACURE***HemaSyst***tm***Contents: One ***DUAL CANNULA TIP***WITH ADAPTER***16 ga. x 4" (10.2 cm)***REF H-1604*** CE 0297***LOT 111144***2014/10***40680b/20004-02***CAUTION: U.S. Federal Law restricts this device to sale by or on the order of a licensed physician (or properly licensed practitioner).***Contents of unopened undamaged package are sterile***CE 0297***Manufactured by: ***Micromedics*** St. Paul, MN 55121-1385 USA***Distributed by:***Haemacure Corporation***Sarasota, FL 34236 USA***Toll Free 1-877-872-4583***www.haemacure.com***Box label***Contents: 5 packages of 1 each***DUAL CANNULA TIP***WITH ADAPTER***16 ga. x 4" (10.2 cm)***REF H-1604***CE 0297***STERILE EO***LOT 11144***2014/10***40681B/20544***". Sterile, single-use instrument for the application of two liquids.
Codes / lots: Model H-1604, Lot Number 111144, expiration 10/2014
Reason for recall
This product has been recalled due to the lot number being different on the box and pouch. The lot number on the box label was incorrectly printed as 11144 instead of 111144. The pouch label referenced the correct lot number, 111144.
FDA-determined cause: Labeling mix-ups
Action taken
Haemacure Corporation, Sarasota, FL initiated letters via e-mail on 11/16/2009 with delivery and read receipt confirmation to their two consignees. The letters identified the affected product and stated the reason for the recall. The letter also states that the firm will replace each order upon receipt of the mislabeled boxes at no charge to the customer. Customers were asked to return their unused devices. Questions should be directed to customer service at 1-877-872-4583.
Details
- Recall number
- Z-1537-2012
- Initiated
- November 16, 2009
- Posted by FDA
- May 11, 2012
- Terminated
- May 14, 2012
- Product code
- KYZ – Syringe, Irrigating (Non Dental)
- Quantity
- 7 boxes
- Distribution
- Nationwide Distribution -- Illinois and California.
- Recalling firm
- Haemacure Corporation, Sarasota, FL