Recall Z-1545-2016

Smith & Nephew, Inc. · initiated March 11, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

TC-PLUS(TM), Primary Fixed Tibial Component VKS Left, SIZE 6 LEFT, STERILE R, REF 12 000 018, S&N 7500273, Smith & Nephew Orthopaedics AG Product Usage: Tibial knee prosthesis

Codes / lots: Batch numbers: B1510531, B1506602, B1502407

Reason for recall

The Affected Devices may contain a dimensional deviation that could potentially result in the surgeon being unable to seat the polyethylene (PE) Insert on the affected VKS tibial base plate during surgery.

FDA-determined cause: Process control

Action taken

The firm, Smith & Nephew, issued an "URGENT FIELD SAFETY NOTICE: Medical Device Field Safety Corrective Action/Recall" letter dated 3/10/2016 to its local QA/RA personnel/customers in the affected countries by email on 03/11/2016. The letter described the product, problem and actions to be taken. The customers were instructed to locate and quarantine affected unused devices immediately; return quarantined product to your national Smith & Nephew agency/distributor; complete and return the Return Slip via fax to your national Smith & Nephew agency/distributor; and pass on this safety information to all those who need to be aware in your organization. The local QA/RA staff will then notify end customers according to local procedures. In Germany as well as in Switzerland, sales representatives will inform the affected customers in-person and retrieve affected devices from their stock. If you have any questions, call 978-749-1330 or email: joe.metzger@smith-nephew.com

Details

Recall number
Z-1545-2016
Initiated
March 11, 2016
Posted by FDA
April 21, 2016
Terminated
April 10, 2017
Product code
JWH – Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Quantity
22 units
Distribution
International Distribution only in countries of: Germany and Switzerland.
Recalling firm
Smith & Nephew, Inc., Memphis, TN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1537-2016 Class IITC-PLUS(TM), Primary Fixed Tibial Component VKS Left, SIZE 2 LEFT, STERILE R, REF 12 000…The Affected Devices may contain a dimensional deviation that could potentially result in the surgeon being… 03/11/2016Terminated
Z-1538-2016 Class IITC-PLUS(TM), Primary Fixed Tibial Component VKS Left, SIZE 8 LEFT, STERILE R, REF 12 000…The Affected Devices may contain a dimensional deviation that could potentially result in the surgeon being… 03/11/2016Terminated
Z-1539-2016 Class IITC-PLUS(TM), Primary Fixed Tibial Component VKS Left, SIZE 10 LEFT, STERILE R, REF 12…The Affected Devices may contain a dimensional deviation that could potentially result in the surgeon being… 03/11/2016Terminated
Z-1540-2016 Class IITC-PLUS(TM), Primary Fixed Tibial Component VKS Right, SIZE 2 RIGHT, STERILE R, REF 12…The Affected Devices may contain a dimensional deviation that could potentially result in the surgeon being… 03/11/2016Terminated
Z-1541-2016 Class IITC-PLUS(TM), Primary Fixed Tibial Component VKS Right, SIZE 6 RIGHT, STERILE R, REF 12…The Affected Devices may contain a dimensional deviation that could potentially result in the surgeon being… 03/11/2016Terminated
Z-1542-2016 Class IITC-PLUS(TM), Primary Fixed Tibial Component VKS Right, SIZE 8 RIGHT, STERILE R, REF 12…The Affected Devices may contain a dimensional deviation that could potentially result in the surgeon being… 03/11/2016Terminated
Z-1543-2016 Class IITC-PLUS(TM), Primary Fixed Tibial Component VKS Right, SIZE 10 RIGHT, STERILE R, REF 12…The Affected Devices may contain a dimensional deviation that could potentially result in the surgeon being… 03/11/2016Terminated
Z-1544-2016 Class IITC-PLUS(TM), Primary Fixed Tibial Component VKS Left, SIZE 4 LEFT, STERILE R, REF 12 000…The Affected Devices may contain a dimensional deviation that could potentially result in the surgeon being… 03/11/2016Terminated
Z-1546-2016 Class IITC-PLUS(TM), Primary Fixed Tibial Component VKS Left, SIZE 8 LEFT, STERILE R, REF 12 000…The Affected Devices may contain a dimensional deviation that could potentially result in the surgeon being… 03/11/2016Terminated
Z-1547-2016 Class IITC-PLUS(TM), Primary Fixed Tibial Component VKS Right, SIZE 4 RIGHT, STERILE R, REF 12…The Affected Devices may contain a dimensional deviation that could potentially result in the surgeon being… 03/11/2016Terminated
Z-1548-2016 Class IITC-PLUS(TM), Primary Fixed Tibial Component VKS Right, SIZE 6 RIGHT, STERILE R, REF 12…The Affected Devices may contain a dimensional deviation that could potentially result in the surgeon being… 03/11/2016Terminated
Z-1549-2016 Class IITC-PLUS(TM), Primary Fixed Tibial Component VKS Right, SIZE 8 RIGHT, STERILE R, REF 12…The Affected Devices may contain a dimensional deviation that could potentially result in the surgeon being… 03/11/2016Terminated
Z-1550-2016 Class IITC-PLUS(TM), Primary Fixed Tibial Component VKS Right, SIZE 10 RIGHT, STERILE R, REF 12…The Affected Devices may contain a dimensional deviation that could potentially result in the surgeon being… 03/11/2016Terminated
Z-1551-2016 Class IITC-PLUS(TM), Primary Fixed Tibial Component VKS Right, SIZE 12 RIGHT, STERILE R, REF 12…The Affected Devices may contain a dimensional deviation that could potentially result in the surgeon being… 03/11/2016Terminated