Recall Z-1556-2020

Abbott Diabetes Care, Inc. · initiated November 14, 2019

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Precision Xtra Blood Glucose & Ketone Monitoring System, UDI: 00093815715023, 00093815988144

Codes / lots: Not associated with specific model/lot numbers.

Reason for recall

Instructions sent in a letter or in instructions for use for disinfecting blood glucose and/or ketone meters were distributed but not cleared by the U.S. FDA, and devices were not cleared for multiple-patient use. Instructions could lead to inappropriate meter disinfection or improper handling of contaminated meters, which may cause infection or exacerbation of an existing infection.

FDA-determined cause: Labeling False and Misleading

Action taken

On 11/14/19, Urgent Medical Device Correction notices were mailed to customers via overnight delivery. Customers were told to continue to use instructions for use regarding the cleaning of meters for use on single patients only. Customers were further asked to complete and return customer response forms. Customers with additional questions were encouraged to call customer service in 1-888-522-5226.

Details

Recall number
Z-1556-2020
Initiated
November 14, 2019
Posted by FDA
February 28, 2020
Terminated
April 28, 2021
Product code
NBW – System, Test, Blood Glucose, Over The Counter
Distribution
U.S.: KY, NY, TN, MA, IL, SC, VT, FL, CA, NC, MS, NJ, TX, AL, GA, OK, CT, PA, NE, ND, MT, MO, DC, ME, LA, KS, ID, OH, MN, NM, VA, IN
Recalling firm
Abbott Diabetes Care, Inc., Alameda, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1557-2020 Class IIOptium EZ Blood Glucose Monitoring System, UDI: 00093815710653Instructions sent in a letter or in instructions for use for disinfecting blood glucose and/or ketone meters… 11/14/2019Terminated
Z-1558-2020 Class IIFreeStyle Precision H Blood Glucose and Ketone Monitoring System, UDI: 00093815711810Instructions sent in a letter or in instructions for use for disinfecting blood glucose and/or ketone meters… 11/14/2019Terminated