Recall Z-1558-2013

Verathon, Inc. · initiated May 8, 2013

Class I Terminated

Class I – dangerous or defective products that could cause serious health problems or death

Product

GlideScope Video Laryngoscope (GVL) 3, 4, 5 and GlideScope (AVL) 2, 3, 4 and 5 reusable blades Product Usage: The GlideScope Video Laryngoscopes are intended for use by qualified medical professionals to obtain a clear, unobstructed view of the vocal cords for medical procedures. The GlideScope Video Laryngoscopes system should only be used by individuals who have been trained and authorized by a physician or the institution providing patient care. The device is not life sustaining or life supporting. The GlideScope AVL Video Monitor system has the same intended use as the GVL System. The difference between systems is that the AVL has a digital platform and the GVL an analog platform.

Codes / lots: Model, Part and Serial Numbers: Model # GVL 3, Part # 0574-0007, Serial # range MD112388 to MD121908; Model # GVL 4 Part # 0574-0001 Serial # range LG112759 to LG122582; Model # GVL 5 Part # 0574-0030 Serial # range XL111799 to XL121759; Model # AVL 2 Part # 0574-0118 Serial # range AC111500 to AC121604; Model # AVL 3 Part # 0574-0115 Serial # range AD111500 to AD121688; Model # AVL 4 Part # 0574-0116 Serial # range AE11150 to AE121778; and Model # AVL 5 Part # 0574-0117 Serial # range AF111500 to AF121666.

Reason for recall

Verathon GlideScope Reusable GVL and AVL Blades are being recalled due to the potential risk of premature failure/breakage of the blade tip that may not be readily visible during routine inspection before or after intubation.

FDA-determined cause: Device Design

Action taken

Verathon issued an Urgent - Safety Alert and Medical Device Recall letter, dated May 10, 2013, to all affected customers The letter identified the affected product, problem and actions to be taken. The letter advises customers to discontinue the use of the specific serial numbers listed in the letter and asked that the affected product be returned to Verathon for replacement. The customers are also directed to fill out a return response form (included in the letter) regarding this recall. Consignees with questions are directed to contact their Verathon representative or Verathon Customer Care at 800-331-2313 (US and Canada) and 1-425-867-1348 for those consignees outside of US and Canada. the support team will be available Monday through Friday from 6 AM - 5 PM PST for the US and from 6 AM to 4:30 pm PST outside the US. Alternately they can e-mail Verathon at customerservice@verathon.com

Details

Recall number
Z-1558-2013
Initiated
May 8, 2013
Posted by FDA
June 20, 2013
Terminated
December 17, 2013
Product code
CCW – Laryngoscope, Rigid
Quantity
4784 units worldwide (3,381 in the US and 1403 international)
Distribution
Worldwide distribution: USA (nationwide) including states of : AK, AL, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, Washington DC, Guam, and countries of : Argentina, Australia, Brazil, Canada, China, Ecuador, France, Germany, Hong Kong, India, Indonesia, Italy…
Recalling firm
Verathon, Inc., Bothell, WA

Official FDA recall record