Recall Z-1560-2010

Animas Corporation · initiated December 18, 2009

Terminated

Product

One Touch Ping Insulin Pump using ezManager Max diabetes management software

Codes / lots: Pump serial numbers: 00-72187-15, 00-72194-15, 01-72188-15, 02-72196-15, 04-70673-15, 05-72199-15, 11-70768-15, 18-67503-15, 23-69304-15, 25-66538-15, 30-69170-15, 40-69613-15, 67-69696-15, 75-72201-15, 76-72202-15, 89-71176-15, 90-71177-15, 96-70468-15, 97-72191-15, 98-72192-15, and 99-72193-15.

Reason for recall

Software malfunction which disallows users to download, view and print information from their pump (including blood glucose, insulin delivery, carbohydrate and pump events). Additionally, users are unable to customize the built-in food database and pump settings using the software.

FDA-determined cause: Software design

Action taken

The recalling firm telephoned the patients/users to inform them of the problem. A letter dated 12/22/09 was also sent to users. Users will have a replacement unit sent to them so that they will be able to fully utilize the pump and the software. As a courtesy a letter dated 12/22/09 was issued to the health care provider/prescribing physician for the affected users.

Details

Recall number
Z-1560-2010
Initiated
December 18, 2009
Posted by FDA
May 8, 2010
Terminated
June 22, 2010
Product code
LZG – Pump, Infusion, Insulin
Quantity
21 pumps
Distribution
The pumps were provided to patients/users in CA, CT, FL, GA, ID, MA, MO, NJ, OH, OR, PA, TX, and WA
Recalling firm
Animas Corporation, West Chester, PA