Recall Z-1560-2014

Teleflex Medical · initiated March 11, 2014

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Deknatel Teleflex Medical Stainless Steel 2 (metric 5) Monofilament Stainless Steel Suture, Nonabsorbable Surgical Suture, U.S.P. Used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Codes / lots: Product Code X-4981M4 - Lot 02J0800451

Reason for recall

Product does not meet minimum needle attachment strength requirements; thus, the needle can become detached from the suture during use.

FDA-determined cause: Employee error

Action taken

Urgent Medical Device Recall Notification letters were issued on March 11, 2014 to consignees. The letter requested that customers discontinue use and quarantine any product on hand. The attached Recall Acknowledgment Form should be completed and returned to Teleflex regardless of whether affected product is on hand or not. Once received a customer service representative will contact customers with instructions on how to return the affected product. Questions should be directed to Customer Service at 1-866-246-6990.

Details

Recall number
Z-1560-2014
Initiated
March 11, 2014
Posted by FDA
May 7, 2014
Terminated
August 14, 2017
Product code
GAQ – Suture, Nonabsorbable, Steel, Monofilament And Multifilament, Sterile
Quantity
32,271 ea.
Distribution
Worldwide Distribution -- USA, including the states of AL, CA, FL, GA, IL, IN, KS, LA, ME, MA, MI, MN, MS, MO, NV, NJ, NM, NY, OH, PA, RI, SC, TN, TX, UT, VA, WV, and WI; and countries of Canada, Germany, and Ireland.
Recalling firm
Teleflex Medical, Durham, NC

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1555-2014 Class IINonabsorbable Surgical Suture, U.S.P., Sterile - Do Not Resterilize, Deknatel. Indicated…Product does not meet minimum needle attachment strength requirements; thus, the needle can become detached… 03/11/2014Terminated
Z-1556-2014 Class IITeleflex Medical, Research Triangle Park, NC, Braided Polyester Rx Only, Sterile…Product does not meet minimum needle attachment strength requirements; thus, the needle can become detached… 03/11/2014Terminated
Z-1557-2014 Class IITeleflex Medical, Research Triangle Park, NC, Polypropylene, Rx Only, Sterile…Product does not meet minimum needle attachment strength requirements; thus, the needle can become detached… 03/11/2014Terminated
Z-1558-2014 Class IIFixt Suture Bonded Plus Violet Braided Polyglycolic Acid Coated Suture, Synthetic…Product does not meet minimum needle attachment strength requirements; thus, the needle can become detached… 03/11/2014Terminated
Z-1559-2014 Class IIBard Fixt Suture Braided Polyester, Rx Only, Teleflex Medical, Research Triangle Park…Product does not meet minimum needle attachment strength requirements; thus, the needle can become detached… 03/11/2014Terminated
Z-1561-2014 Class IIDeknatel Teleflex Medical Silk Black Braided Silk Suture Nonabsorbable Surgical Suture…Product does not meet minimum needle attachment strength requirements; thus, the needle can become detached… 03/11/2014Terminated