Recall Z-1569-2013

Terumo Cardiovascular Systems Corporation · initiated May 30, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Sarns " TCM II Temperature Control Module The Sarns" TCM II Temperature Control Module is a source of temperature-controlled water for blood heat exchangers used in an extracorporeal circuit and is indicated for controlling and monitoring patient temperature

Codes / lots: Catalog Number 4416 and Serial Numbers: 4656, 4657, 4658,4659, 4661, 4662, and 4663.

Reason for recall

Internal testing found certain Sarns" TCM II systems were released with incorrectly assembled heatsink assemblies which could cause the units to fail.

FDA-determined cause: Employee error

Action taken

Terumo sent beginning May 30, 2013, an URGENT MEDICAL DEVICE RECALL NOTIFICATION were sent to all customers with instructions to notify all customers of the recall requesting all affected units to be quarantined until the heatsink units are replaced Confirm receipt by filling out the response form containing communication by faxing or emailing the attached Affiliate Response Form to the fax number or email address indicated on the form. Questions should be directed to Terumo CVS customer service in the U.S.: Telephone: +1-734-663-4145; Fax: +1-734-741-6149

Details

Recall number
Z-1569-2013
Initiated
May 30, 2013
Posted by FDA
June 21, 2013
Terminated
September 17, 2013
Product code
DWC – Controller, Temperature, Cardiopulmonary Bypass
Quantity
7 units
Distribution
Worldwide Distribution including the countries of UAE (United Arab Emirates) HONG KONG and THAILAND. Not distributed in USA
Recalling firm
Terumo Cardiovascular Systems Corporation, Ann Arbor, MI