Recall Z-1572-2025

Abbott Diagnostics Scarborough, Inc. · initiated March 4, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Brand Name: ID NOW Influenza A/B 2 24T Product Name: ID NOW Influenza A/B 2 24T Model/Catalog Number: 427-000 Software Version: N/A Product Description: Influenza A And Influenza B Multiplex Nucleic Acid Assay; Real Time Nucleic Acid Component: No

Codes / lots: Model 427-000; GTIN/DI 10811877010422; Lot 000M889669 Model 427-000; GTIN/DI 10811877010422; Lot 000M890889 Model 427-000; GTIN/DI 10811877010422; Lot 000M916327 Model 427-000; GTIN/DI 10811877010422; Lot 000M916649 Model 427-000; GTIN/DI 10811877010422; Lot 000M916672 Model 427-000; GTIN/DI 10811877010422; Lot 000M916724 Model 427-000; GTIN/DI 10811877010422; Lot 000M917131 Model 427-000; GTIN/DI 10811877010422; Lot 000M917164 Model 427-000; GTIN/DI 10811877010422; Lot 000M917180 Model 427-000; GTIN/DI 10811877010422; Lot 000M917508 Model 427-000; GTIN/DI 10811877010422; Lot 000M917581 Model 427-000; GTIN/DI 10811877010422; Lot 000M919907 Model 427-000; GTIN/DI 10811877010422; Lot 000M921626 Model 427-000; GTIN/DI 10811877010422; Lot 000M921636 Model 427-000; GTIN/DI 10811877010422; Lot 000M921728 Model 427-000; GTIN/DI 10811877010422; Lot 000M923485 Model 427-000; GTIN/DI 10811877010422; Lot 000M926234 Model 427-000; GTIN/DI 10811877010422; Lot 000M926370 Model 427-000; GTIN/DI 10811877010422; Lot 000M927991 Model 427-000; GTIN/DI 10811877010422; Lot 000M928421 Model 427-000; GTIN/DI 10811877010422; Lot 000M929248 Model 427-000; GTIN/DI 10811877010422; Lot 000M931072 Model…

Reason for recall

The impacted lots have a higher occurrence of invalid rates when compared to the product Instructions for Use.

FDA-determined cause: Process change control

Action taken

On March 4, 2025 FIELD CORRECTION NOTICE letters were sent to end user customers. Please complete the following actions, as applicable. 1. If you have impacted inventory in stock, then: " Discontinue use of and destroy any remaining inventory of the impacted lots according to your procedures. " Complete and return the Customer Reply Form (Form must be completed, signed and returned to receive a replacement lot). " Please retain this letter for your records. 2. If you have forwarded the product listed above to others in your network, then: - Inform them of this Field Correction Notice, provide to them a copy of this notice and request they take the necessary action. 3. If you do not have impacted inventory in stock, then: - All product lots not identified in Appendix I can continue to be used. If you have questions regarding this information, please contact Sedgwick by phone at 888 671 8858 (available 8am 5pm EST, Monday - Friday). Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online (http://www.fda.gov/MedWatch/report.htm), by mail (http://www.fda.gov/MedWatch/getforms.htm), by phone (1-800-332-1088), or by fax (1-800-FDA-0178). If you have experienced any patient or user injury associated with this Field Correction Notice, please immediately report the event to your local area Customer Service. ****Update added 5/8/2025**** On April 29, 2025 FIELD CORRECTION NOTICE letters were sent to customers informing them of additional lots included in this recall.

Details

Recall number
Z-1572-2025
Initiated
March 4, 2025
Posted by FDA
April 15, 2025
Product code
OZE – Influenza A And Influenza B Multiplex Nucleic Acid Assay
Quantity
592320 eaches (535416 US, 56904 OUS)
Distribution
Worldwide distribution - US Nationwide and the countries of Australia; Austria; Belgium; Bulgaria; Canada; Costa Rica; France; French Polynesia; Germany; Italy; Luxembourg; Malaysia; Netherlands; Paraguay; Poland; Qatar; Saudi Arabia; Slovenia; South Korea; Spain; Switzerland; Taiwan; United Kingdom; Uruguay.
Recalling firm
Abbott Diagnostics Scarborough, Inc., Scarborough, ME

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1571-2025 Class IIBrand Name: ID NOW COVID-19 2.0 24T Product Name: ID NOW COVID-19 2.0 24T Model/Catalog…The impacted lots have a higher occurrence of invalid rates when compared to the product Instructions for Use. 03/04/2025Open, Classified
Z-1573-2025 Class IIBrand Name: ID NOW RSV 24T Product Name: ID NOW RSV 24T Model/Catalog Number: 435-000…The impacted lots have a higher occurrence of invalid rates when compared to the product Instructions for Use. 03/04/2025Open, Classified