Recall Z-1577-2025

Maquet Cardiovascular, LLC · initiated March 10, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Maquet VH-3010 Power Supply. Used to deliver power to the harvesting tool of the Vasoview Hemopro EVH System (HemoPro 2).

Codes / lots: Model No. VH-3010; UDI-DI: 0060756770826; Serial No. H24080065G, H24080074G, H24080012G, H24080022G, H24080184G, H24080185G, H24080051G, H24080117G, H24080069G, H24080016G, H24080079G, H24080041G, H24080056G, H24080030G, H24080068G, H24080034G, H24080058G, H24080026G, H24080033G, H24080020G, H24080032G, H24080053G, H24080066G, H24080077G, H24080144G, H24080123G, H24080042G, H24080070G, H24080008G, H24080003G, H24080174G, H24080064G, H24080027G, H24080044G, H24080047G, H24080049G, H24080061G, H24080078G, H24080010G, H24080138G, H24080080G, H24080006G, H24080015G, H24080038G, H24080085G, H24080084G, H24080031G, H24080101G, H24080148G, H24080040G, H24080055G, H24080025G, H24080052G, H24080048G, H24080093G, H24080126G, H24080073G, H24080007G, H23060082G, H24080075G, H24080113G, H24080127G, H24080011G, H24080005G, H24080013G, H24080023G, H24080057G, H24080110G, H24080009G, H24080067G, H24080029G, H24080043G, H24080183G, H24080165G, H24080017G, H24080001G, H24080037G, H24080060G, H24080072G, H24080076G, H24080045G, H24080024G, H24080028G, H24080114G, H24080136G, H24080062G, H24080018G, H24080054G, H24080063G, H24080112G, H24080118G, H24080071G, H24080182G, H24080002G, H24080004G…

Reason for recall

Incorrect resistor utilized in the VH-3010 Power Supply, which may cause the jaws of the HemoPro 2 to not adequately heat up.

FDA-determined cause: Under Investigation by firm

Action taken

An URGENT MEDICAL DEVICE REMOVAL notification dated March 2025 was mailed to consignees. The notification instructs consignees to examine their inventory for impacted serial numbers. For power supplies whose serial numbers are not listed in the provided Appendix A of recalled devices, devices can continue to be used. If power supplies with affected serial numbers are identified as being in stock, the associated devices should not be used and returned to Getinge. Returns are facilitated by contacting Getinge Customer Service at 888-800-2874 between the hours of 6:00 AM and 5:00 PM PST to request a return authorization number and shipping instructions. Consignees are to forward the provided notification to all users of the device within the facility and forwarded to any locations where devices were further distributed. Consignees with any questions about the recall are to contact their Getinge representative.

Details

Recall number
Z-1577-2025
Initiated
March 10, 2025
Posted by FDA
April 16, 2025
Product code
GEI – Electrosurgical, Cutting & Coagulation & Accessories
Quantity
110 units
Distribution
Worldwide - US Nationwide distribution in the states of AL, CA, CO, DC, FL, GA, IA, ID, IL, IN, KY, MD, MI, MN, MO, NC, NH, NJ, NY, OH, PA, PR, TX, UT, VA, WI, WV and the countries of China, Japan, Netherlands, New Zealand, Singapore, United Kingdom, United States.
Recalling firm
Maquet Cardiovascular, LLC, Wayne, NJ