Recall Z-1586-2013

Medrad Inc · initiated April 15, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

DC power cable (p/n 3010556) supplied with the MEDRAD(R) Veris MR Vital Signs Monitor: Product Usage: Monitor, Physiological, Patient

Codes / lots: Catalog numbers 3011992, 3011992W, 3011993, 3011993W, 3011994, 3011994W, 3011996, 3011996W, 3014268 and 3014268W.

Reason for recall

The firm is recalling these power cables due to a latent design reliability issue and the potential for shorting which can result in heating/melting of the cable jacket.

FDA-determined cause: Device Design

Action taken

Medrad Inc sent an Urgent Medical Device Component Recall letter dated April 15, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter states that replacement products will be available in June and shipped in the coming months along with additional instructions for returning the affected product. Customers were instructed to complete the enclosed recall response form and fax it back to Bayer at 412-406-0942. For questions contact customer support team at 1-877-229-3767 and select option 2.

Details

Recall number
Z-1586-2013
Initiated
April 15, 2013
Posted by FDA
June 25, 2013
Terminated
April 3, 2014
Product code
MWI – Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)
Quantity
1,066
Distribution
Worldwide Distribution - US Nationwide including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY and the countries of: AE, AU, AZ, BN, BR, CA, CH, CL, CO, DE, DK, DO, DZ, EG, ES, FR, GB, HK, HN, HR, HU, IN, IT, JO, JP, KR, LB, MU, MX…
Recalling firm
Medrad Inc, Indianola, PA