Recall Z-1591-2009

Roche Diagnostics Corp. · initiated May 13, 2009

Terminated

Product

Roche ACCU-Chek Active Meter Kit, Roche Diagnostics, Indianapolis, IN; REF 03184501001 includes the Roche ACCU-CHEK Softclix Lancing Device, Distributed by Roche Diagnostics, Indianapolis, IN. Model 3144844001 (10 lancets). The device is used to prick the finger and to obtain a drop of blood in preparation for diabetes/insulin level testing.

Codes / lots: REF 03184501001; lots 116099, 116109, 116122, 116135, 116145, 116146, 116155, 116167, 116185, 116190, 116192, 116205, 116209, 116215, 116229, 116230 and 116237. Lancet Model 3144844001; lots M25C2, M25C5, M25C7, N32A5 and M32B5.

Reason for recall

The lancet cap may be missing, which could result in an unintended lancet stick to the user.

FDA-determined cause: Other

Action taken

An "Urgent Medical Device Correction" letter dated May 13, 2009 was sent to all affected customers, distributors and consumer consignees who have either registered their blood glucose meter, contacted the firm about the lancets since 10/1/08, or have been shipped lancets by Roche since 10/1/08 were notified of the problem. Distributors were requested to notify each patient self-tester to whom a meter kit was distributed since October 1, 2008. The firm issued a press release on May 18, 2009. Diabetes Care customers: If you have questions regarding the recall, please contact the ACCU-CHEK Softclix Lancet Hotline at 1-800-778-7057. CoaguChek meter customers: If you have questions regarding the recall, please contact the ACCU-CHEK Softclix Lancet Hotline at 1-800-778-7505.

Details

Recall number
Z-1591-2009
Initiated
May 13, 2009
Posted by FDA
August 11, 2009
Terminated
January 5, 2011
Product code
LFR – Glucose Dehydrogenase, Glucose
Quantity
449,533 packages of lancets
Distribution
Nationwide.
Recalling firm
Roche Diagnostics Corp., Indianapolis, IN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1590-2009 Roche ACCU-CHEK Softclix Lancing Device, Distributed by Roche Diagnostics, Indianapolis…The lancet cap may be missing, which could result in an unintended lancet stick to the user. 05/13/2009Terminated
Z-1592-2009 Roche ACCU-Chek Voicemate Meter, Roche Diagnostics, Indianapolis, IN; REF 12030802001…The lancet cap may be missing, which could result in an unintended lancet stick to the user. 05/13/2009Terminated
Z-1593-2009 Roche ACCU-Chek Aviva Meter Kit, Roche Diagnostics, Indianapolis, IN; REF 04893247001…The lancet cap may be missing, which could result in an unintended lancet stick to the user. 05/13/2009Terminated
Z-1594-2009 Roche Coaguchex XS Systems for Patient Self-Testing, Roche Diagnostics, Indianapolis…The lancet cap may be missing, which could result in an unintended lancet stick to the user. 05/13/2009Terminated
Z-1595-2009 Roche ACCU-CHEK Compact Plus Meter Kit, Roche Diagnostics, Indianapolis, IN; REFS…The lancet cap may be missing, which could result in an unintended lancet stick to the user. 05/13/2009Terminated