Recall Z-1591-2026

Abiomed, Inc. · initiated February 16, 2026

Class I Open, Classified

Class I – dangerous or defective products that could cause serious health problems or death

Product

Automated Impella Controller (AIC), labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 5. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 6. Impella Optical Controller, Packaged, UK; Product Code: 0042-0010-UK. 7. Optical AIC w/Impella Connect, Pack'd, CA; Product Code: 0042-0040-CA. 8. Optical AIC w/Impella Connect, Pack'd, EU; Product Code: 0042-0040-EU. 9. Optical, AIC, Impella Connect, Pkgd, AU; Product Code: 0042-0040-AU.

Codes / lots: All products below with software version 8.2.2 through 8.5 are affected. 1. Product Code: 0042-0000-CA; GTIN: 00813502011272; Serial Numbers: IC2035, IC9416. 2. Product Code: 0042-0000-EU; GTIN: 00813502011289; Serial Numbers: IC1677, IC3049, IC3099, IC3102, IC4358, IC4329, IC5774, IC5757, IC6580, IC6581. 3. Product Code: 0042-0000-UK; GTIN: 00813502011296; Serial Numbers: IC2023, IC1910, IC2704, IC3325, IC3312. 4. Product Code: 0042-0000-US; GTIN: 00813502010022; Serial Numbers: IC1140. 5. Product Code: 0042-0010-EU; GTIN: 00813502010961; Serial Numbers: IC5403, IC5404, IC5635, IC5636, IC5666, IC5667, IC5801, IC5904, IC5918, IC5921, IC6421, IC6422, IC8683. 6. Product Code: 0042-0010-UK; GTIN: 00813502010978; Serial Numbers: IC8101, IC8102, IC8103, IC8104, IC8544, IC9446, IC9449, IC11440. 7. Product Code: 0042-0040-CA; GTIN: 00813502012958; Serial Numbers: IC10306, IC10312, IC12604, IC12891, IC12895, IC12892, IC12605, IC12894. 8. Product Code: 0042-0040-EU; GTIN: 00813502012217; Serial Numbers: IC9014, IC9022, IC9429, IC10408, IC10425. 9. Product Code: 0042-0040-AU; GTIN: 00813502012972; Serial Numbers: IC10163, IC10387, IC11546, IC11566, IC11490, IC11544, IC12008, IC12054…

Reason for recall

Potential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps.

FDA-determined cause: Software design

Action taken

Abiomed notified international consignees on about 02/16/2026 via emailed CRITICAL PRODUCT CORRECTION letter. Consignees were instructed to cease use of the Automated Impella Controller (AIC) with an Impella 5.5 S1 (product code 0550-0007) unless AIC software is updated to v10.1 or above, identify current software in the lower left corner of the initial AIC display screen, contact your local Abiomed representative to schedule your software update, review and complete the customer response form provided and post this notification in a visible area for awareness. Consignees were also instructed to forward this notification to anyone within the facility that should be aware and forward the notification to consignees if further distributed.

Details

Recall number
Z-1591-2026
Initiated
February 16, 2026
Posted by FDA
March 27, 2026
Product code
OZD – Temporary Non-Roller Type Left Heart Support Blood Pump
Quantity
62 units
Distribution
International distribution in the country of Australia, Canada, France, Germany, Kuwait.
Recalling firm
Abiomed, Inc., Danvers, MA

Official FDA recall record