Recall Z-1594-2024
Randox Laboratories Ltd. · initiated January 31, 2024
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Microalbumin Calibrator Series (mALB CAL)-IVD to calibrate the Microalbumin Assay Catalog Number: MA2426
Codes / lots: GTIN: 05055273204544 (1) Batch/Lot: 602114 /1083MA-1087MA (2) Batch/Lot: 636645 /1108MA-1112MA (3) Batch/Lot: 662429 /1083MA- 1087MA
Reason for recall
Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other method may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%.
FDA-determined cause: Under Investigation by firm
Action taken
Randox Laboratories (Manufacturer) issued Urgent Field Safety Notice on 31 Jan 24 Complaint Reference: REC716 letter via email to Distributors. Distributors will contact the end-users. Letter states reason for recall, health risk and action to take: We have reassigned the target values in the above lot numbers in line with the ERM reference material. Please refer to the table below for the updated calibrator targets. You may experience a shift in Quality Control and patient sample recovery of up to +12%. Please discard all copies of the calibrator Instructions For Use (IFU)s and download the updated IFUs from www.randox.com . If further information is required, please contact technical.services@randox.com . In line with the restandardisation to the Microalbumin Calibrator Series, we are updating the target and range for the Microalbumin Liquid control, MA1361, lots 1102MA and 1103MA which are present in batch 631698. The updated targets and ranges can be found in the table below. Please discard any copies of the IFU and download the updated version from www.randox.com.
Details
- Recall number
- Z-1594-2024
- Initiated
- January 31, 2024
- Posted by FDA
- April 17, 2024
- Product code
- DDZ – Albumin, Fitc, Antigen, Antiserum, Control
- Quantity
- 6 units
- Distribution
- US Nationwide distribution in the states of GA, MA, WV including PR.
- Recalling firm
- Randox Laboratories Ltd., Crumlin, North
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1593-2024 Class II | Microalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay…Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other methods… | 01/31/2024Open, Classified |