Recall Z-1605-2024

Seaspine Orthopedics Corporation · initiated February 15, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Mariner RDX Polyaxial Head, REF: MX1-000020; Mariner RDX Extended Polyaxial Head, REF: MX1-000030; Mariner RDX Deformity Head, REF: MX1-000040; Mariner RDX Extended Deformity Head, REF: MX1-000050; Mariner RDX Trauma Head, REF: MX1-000060; Mariner RDX Extended Trauma Head. REF: MX1-000070, part of Mariner RDX Alpha Instruments and Implants, REF: AMRDX and Mariner RDX TD Alpha Implants, REF: AMRDXTD, which are pedicle screw systems.

Codes / lots: REF/UDI-DI/Lots: MX1-000020/10889981289751/AW161073C, AW164798D, AW164799D, AW164800E, AW164801E; MX1-000030/10889981289768/AW164802F, AW164803G, AW157588F, AW161074F, AW163331F; MX1-000040/10889981289775/AW157589D, AW164806E; MX1-000050/10889981289782/AW164807G, AW164808G, AW164809H, AW157590G, AW172902H; MX1-000060/10889981289799/AW157591D, AW161076E, AW164811E; MX1-000070/10889981289805/AW163350G, AW164812G, AW157592G, AW161077G; AMRDX/10889981291266/AMRDX-001, AMRDX-002, AMRDX-003, AMRDX-004, AMRDX-005, AMRDX-008, AMRDX-009; AMRDXTD/10889981295332/AMRDXTD-003, AMRDXTD-004, AMRDXTD-005

Reason for recall

Screw heads, of pedicle screw systems, have thin lead-in thread that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.

FDA-determined cause: Nonconforming Material/Component

Action taken

On 2/14/24 the firm phoned distributors and informed them of the recall issue and next steps. On 2/16/24 the firm sent a recall communication via email. Customers were asked to arrange for the return of these sets along with any additional Mariner RDX inventory using a firm provided return label. Customers with questions or concerns can contact the Associate Product Manager, at olivia.disanto@seaspine.com

Details

Recall number
Z-1605-2024
Initiated
February 15, 2024
Posted by FDA
April 22, 2024
Product code
NKB – Thoracolumbosacral Pedicle Screw System
Quantity
1579 Screw Heads
Distribution
US Nationwide distribution in the states of MI, GA, VA, IN, FL.
Recalling firm
Seaspine Orthopedics Corporation, Carlsbad, CA

Official FDA recall record