Recall Z-1606-2018

Philips Medical Systems Nederlands · initiated March 26, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Mutlidiagnost-Eleva with Flat Detector 708034 As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.

Codes / lots: System Serial # 201 137 223 228 219 60 272 35 111 85 121 38 120 185 168 32 30 100 235 34 148 149 67 269 191 118 170 241 210 255 8 273 33 240 131 99 193 36 138 96 222 278 257 246 23 147 162 220 47 179 221 126 195 46 144 254 189 75 56 217 83 5 167 6 154 155 156 177 183 157 245 116 62 207 81 113 134 103 115 55 200 180 175 78 93 24 106 110 123 266 3 URF.173 14 202 208 247 225 28 216 262 9 176 261 259 174 114 68 63 122 135 205 224 39 98 80 22 251 142 253 84 181 264 163 73 94 95 160 199 206 159 112 196 151 25 238 150 4 117 50 58 51 79 249 119 204 133 18 13 15 16 17 129 244 227 234 130 31 12 27 11 128 171 237 76 1 143 248 109 42 43 44 172 213 212 2 48 127 19 82 69 70 65 242 54 194 136 190 152 105 233 71 97 132 197 53 211 72 187 209 41 146 125 268 26 77 40

Reason for recall

The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.

FDA-determined cause: Nonconforming Material/Component

Action taken

The firm, BU IGT Systems, sent an "URGENT- Field Safety Notice" letter dated 3/23/2018 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to stop using the system and call your Philips representative if the following situations occurs: -the table height or tilt movement is not working. - you notice a cracking or snapping sound during system operation. - you experience a blockade in the table height or tilt movement during system movements. -the system is having an unexpected collision or the system had an unexpected collision in the past 2 months. - you have used a CPR stand that obstructed the system, or - you notice any (other) unusual system behavior beyond the normal use of the system. The customers should also avoid collisions with the table and base stand of the system by ensuring that no obstacles are placed around the system. In addition, ensure that all staff with access to the affected systems are informed of the content of this Field Safety Notice. The firm stated: -All possibly affected products in the field will be corrected by means of a field change order free of charge. -A firm representative will replace affected material on the upper and lower tilt actuator in the affected systems. -You will be contacted by our local Philips representative to schedule this corrective action. -This action will start effective April 2018. If you need any further information or support concerning this issue, please contact your local Philips representative: Technical Support Line: 1-800-722- 9377, 8am to 8pm (EST) Monday thru Friday.

Details

Recall number
Z-1606-2018
Initiated
March 26, 2018
Posted by FDA
May 3, 2018
Terminated
June 12, 2020
Product code
KPR – System, X-Ray, Stationary
Quantity
206
Distribution
Worldwide Distribution: US (Nationwide) and countries of: Algeria, Argentina, Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Farce Islands, Finland, France, Germany, Hong Kong, Indonesia, Ireland, Israel, Italy, Jersey, Jordan, Korea, Republic of, Latvia, Lebanon, Luxembourg, Malaysia, Martinique Mexico, Netherlands, New Zealand, Norway, Oman…
Recalling firm
Philips Medical Systems Nederlands, Best

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1603-2018 Class IIMultiDim Eleva 708032 As a multifunctional universal imaging application system, General…The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will… 03/26/2018Terminated
Z-1604-2018 Class II708033 URODiagnost As a multifunctional universal imaging application system, General…The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will… 03/26/2018Terminated
Z-1605-2018 Class IIMultiDiagnost Eleva 708036 As a multifunctional universal imaging application system…The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will… 03/26/2018Terminated
Z-1607-2018 Class IIMutlidiagnost-Eleva with Flat Detector Eleva with Flat Detector 708037 As a…The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will… 03/26/2018Terminated
Z-1608-2018 Class IIMultiDiagnostic-Eleva with Flat Detector 708038 As a multifunctional universal imaging…The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will… 03/26/2018Terminated
Z-1609-2018 Class IIMultiDiagnostic-Eleva with Flat Detector 708035 As a multifunctional universal imaging…The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will… 03/26/2018Terminated