Recall Z-1607-2023
Beckman Coulter Inc. · initiated April 5, 2023
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Beckman Coulter Bicarbonate, REF: OSR6237, 4x50 mL
Codes / lots: All Lots: UDI-GI: 15099590010348
Reason for recall
There is a potential for a decrease in calibration ODs for Bicarbonate reagent which could result in calibration failures and/or quality control failures.
FDA-determined cause: Under Investigation by firm
Action taken
On April 5, 2023, Beckman Coulter (BEC) issued a "URGENT MEDICAL DEVICE RECALL" Notification via mail and/or E-Mail. In addition, to informing consignees about the recalled product, BEC asked consignees to take the following actions: 1. Safely dispose of the Bicarbonate lot numbers that are specifically described in Table A. 2. If replacement or re-imbursement of the above lots is required, customers are directed to contact their local Beckman Coulter representative for replacement and re-imbursement. 3. Customers are advised to implement the adjusted Calibration OD range using the instructions within the customer letter to avoid generation of elevated results for all other Bicarbonate lots. 4. Retrospective review of patient data is not required due to the remote possibility of generating falsely elevated, clinically impactful results which could cause a change in treatment. Please share the content of this letter with your laboratory and/or medical director regarding the need to review previous patient test results. 4. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forward any of the affected products(s) listed above to another laboratory, please provide them a copy of the letter. 5. Please respond within 10 days in one of the following ways: - Electronically, you received this communication via email. - Manually, complete and return the enclosed Response Form. 6. If you have any questions, please call 1-800-854-3633
Details
- Recall number
- Z-1607-2023
- Initiated
- April 5, 2023
- Posted by FDA
- May 18, 2023
- Product code
- KHS – Enzymatic, Carbon-Dioxide
- Quantity
- 48,633 units in total
- Distribution
- Worldwide Distribution: US (nationwide): AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, Guam, & Puerto Rico and OUS (foreign) countries of: Brazil ,Canada, Mexico, Panama, Taiwan,
- Recalling firm
- Beckman Coulter Inc., Brea, CA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1606-2023 Class II | Beckman Coulter Bicarbonate, REF: OSR6137, 4x25 mLThere is a potential for a decrease in calibration ODs for Bicarbonate reagent which could result in… | 04/05/2023Open, Classified |
| Z-1608-2023 Class II | Beckman Coulter Bicarbonate, REF: OS6637, 4x173 mLThere is a potential for a decrease in calibration ODs for Bicarbonate reagent which could result in… | 04/05/2023Open, Classified |