Recall Z-1617-2013

Biosense Webster, Inc. · initiated June 19, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

20-Pole LASSO Nav Catheter Eco, Catalog # D-1349-02S. Designed to facilitate electrophysiological mapping of the atria of the heart.

Codes / lots: Lot # 15832314L

Reason for recall

Biosense Webster is recalling the 20-Pole LASSO NAV Catheter Eco because it has been incorrectly calibrated causing them to be recognized and displayed as a 10-Pole LASSO NAV Catheter Eco by the CARTO 3 System.

FDA-determined cause: Use error

Action taken

An Urgent Field Notice letter was sent to all their customers who purchased the 20-Pole LASSO NAV Catheter Eco. The letter informed the customers of the problem identified and the actions to be taken. Customers are instructed to complete and return the attached Voluntary Field Removal Certification Form in accordance with the instructions listed on the form. Customers are instructed to contact Biosense Webster sales representatives or call (866) 473-7823, Monday-Friday from 7:00AM to 8:00 PM EST if they have any questions. For questions related to the Voluntary Field Removal Certification Form and its return, customers are instructed to contact the Field Action Coordinator, at (909) 839-8468.

Details

Recall number
Z-1617-2013
Initiated
June 19, 2013
Posted by FDA
June 28, 2013
Terminated
February 7, 2014
Product code
DRF – Catheter, Electrode Recording, Or Probe, Electrode Recording
Quantity
10 units
Distribution
Distributed in the states of MN, FL, and UT.
Recalling firm
Biosense Webster, Inc., Irwindale, CA

Official FDA recall record