Recall Z-1620-2011

BD Biosciences, Systems & Reagents · initiated January 3, 2009

Terminated

Product

BD FACS Sample Prep Assistant (SPA and SPA II) Instrument, Model Number: 334080, Product is manufactured and distributed by BD Biosciences, San Jose, CA The intended use for the BD FACS Sample Prep Assistant (SPA and SPA II) is to prepare human whole blood for flow cytometry.

Codes / lots: Lot Number: T0040

Reason for recall

To prevent a needle stick while replacing the probe, the firm has initiated a Field Correction, and issued an improved Users Guide procedure (User Bulletin, Part Number 645050).

FDA-determined cause: Nonconforming Material/Component

Action taken

The firm, BD Biosciences, sent an information letter dated December 2008 with an enclosed New Probe Replacement Service/User Bulletin, to the customers. The letter described the product, problem and action to be taken. The customers were instructed to review the attached User Bulletin (Part Number 645050); ensure that their laboratory follows the updated procedure for removing and replacing probe; place the User Bulletin in their BD FACS SPA or SPA II User's Guide; always follow the instructions in User Bulletin to properly remove and replace probe, and complete and return the Effectivity Form via fax to 408-577-1409. Should you have any comments or questions, please contact BD Biosciences at 1-877-232-8995 in the US. Outside the US, contact your regional BD office or field service engineer for assistance.

Details

Recall number
Z-1620-2011
Initiated
January 3, 2009
Terminated
March 10, 2011
Product code
JQW – Station, Pipetting And Diluting, For Clinical Use
Quantity
70
Distribution
Worldwide distribution: USA and countries including: Hong Kong, Australia, Belgium, Canada, El Salvador, Hong Kong, Japan, and Mexico.
Recalling firm
BD Biosciences, Systems & Reagents, San Jose, CA

Official FDA recall record