Recall Z-1623-2025

Medtronic Xomed, Inc. · initiated March 10, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

NIM Vital Nerve Monitoring System: CONSOLE NIM4CM01 NIM 4.0, REF: NIM4CM01; CONSOLE NIM4CM01RF NIM 4.0 REFURBISHED, REF: NIM4CM01RF; PATIENT INTERFACE NIM4CPB1 NIM 4.0, REF: NIM4CPB1; PATIENT INTFC NIM4CPB1RF NIM 4.0 REFURB, REF: NIM4CPB1RF; SOFTWARE NIM4SWU143 UPGRADE V1.4.3, REF: NIM4SWU143; SOFTWARE NIM4SWU154 UPGRADE V1.5.4, REF: NIM4SWU154; SOFTWARE NIM4SWU164 UPGRADE V1.6.4, REF: NIM4SWU164

Codes / lots: All NIM Vital Nerve Monitoring Systems manufactured or installed with software version v1.6.4 or earlier. REF/UDI-DI: NIM4CM01/763000002978, 00763000395896, 00763000401597, 00763000528577; NIM4CM01RF/00763000002992; NIM4CPB1/763000002985, 00763000401603, 00763000395902, 00763000528584; NIM4CPB1RF/00763000003005; NIM4SWU143/00763000709341, 00763000869823; NIM4SWU154/00763000945398; NIM4SWU164/00763000974312

Reason for recall

Nerve Monitoring System with certain software versions has potential for increased stimulus artifact, which may require troubleshooting, and may lead to procedure delay, extubation, and medical intervention.

FDA-determined cause: Software change control

Action taken

On 3/10/2025, recall notices were mailed, and/or emailed to customers who were asked to do the following: 1) Your Medtronic representative will contact you to install the new software version 1.7.5 for correction of the impacted product in your possession. 2) For patients who are currently being monitored with affected devices, be aware of the possibility of increased stimulus artifact. Refer to the system instructions for use for instructions on how the stimulus artifact may be reduced or exacerbated through the adjustment of system settings including event threshold, stimulation current, and rejection period. 3) Complete and return the customer confirmation form via email to neuro.quality@medtronic.com 4) Share this communication within your organization, with other organizations where impacted devices have been transferred, and any other associated organizations that may be impacted by this action. Maintain a copy of this letter for your records. Distributors are asked to discard any of the items below in your inventory: a. SOFTWARE NIM4SWU143 UPGRADE v1.4.3 b. SOFTWARE NIM4SWU154 UPGRADE v1.5.4 c. SOFTWARE NIM4SWU164 UPGRADE v1.6.4 If you have any questions regarding this communication, please contact the firm's Sales Representative or Technical Services at 1-888-826-5603.

Details

Recall number
Z-1623-2025
Initiated
March 10, 2025
Posted by FDA
April 18, 2025
Product code
GWF – Stimulator, Electrical, Evoked Response
Quantity
9,788
Distribution
Worldwide - US Nationwide distribution including in the states of MD, MN, PA, OK, OH, FL, CO, GA, NC, KS, IL, WY, AK, NE, VA, MI, CA, NY, WI, UT, AZ, AR, TX, WA, AL, NJ, SD, MS, KY, TN, MO, OR, MT, MA, CT, NM, LA, WV, NH, DC, DE, IN, NV, IA, SC, ID, ME, HI, RI, VT, ND, GU, PR and the countries of Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Canary Islands…
Recalling firm
Medtronic Xomed, Inc., Jacksonville, FL

Official FDA recall record