Recall Z-1627-2016

Baxter Healthcare Corp. · initiated April 21, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

GEM(TM), FLOWCOUPLER(R), 2.0 mm, STERILE EO, Rx only, 1 FLOWCOUPLER, REF GEM2752-FC Product Usage: The FlowCOUPLER Device is a single use, implantable device that is intended to be used in the end-to-end anastomosis of veins and arteries normally encountered in microsurgical and vascular reconstructive procedures. The FlowCOUPLER Device includes a pair of permanently implanted rings which secure the anastomosis and a removable Doppler probe that is press-fit onto one of the rings. When the FlowCOUPLER Device is used in conjunction with the FlowCOUPLER Monitor, the FlowCOUPLER System is intended to detect blood flow and confirm vessel patency intra-operatively and post-operatively at the anastomotic site.

Codes / lots: SP15H05-1067843 thru SP15K05 -1094463

Reason for recall

Instructions for use booklet may puncture the outer Tyvek lid.

FDA-determined cause: Package design/selection

Action taken

Baxter sent an Urgent Product Recall letter dated April 21, 2016 to customer. The letter identified the affected product, problem and actions to be taken. The letters requested that direct consignees contact Synovis MCA to arrange for return and credit. Companies that further distributed the product were asked to notify other facilities or departments. For questions contact Synovis MCA at 800-510-3318 or 205-941-0111.

Details

Recall number
Z-1627-2016
Initiated
April 21, 2016
Terminated
November 22, 2016
Product code
DPW – Flowmeter, Blood, Cardiovascular
Quantity
315 units
Distribution
Nationwide and Germany, Sweden, Netherlands, Israel, Denmark, United Kingdom, Canada
Recalling firm
Baxter Healthcare Corp., Deerfield, IL

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1628-2016 Class IIGEM(TM), FLOWCOUPLER(R), 2.5 mm, STERILE EO, Rx only, 1 FLOWCOUPLER, REF GEM2753-FC.…Instructions for use booklet may puncture the outer Tyvek lid. 04/21/2016Terminated
Z-1629-2016 Class IIGEM(TM), FLOWCOUPLER(R), 3.0 mm, STERILE EO, Rx only, 1 FLOWCOUPLER, REF GEM2754-FC…Instructions for use booklet may puncture the outer Tyvek lid. 04/21/2016Terminated
Z-1630-2016 Class IIGEM(TM), FLOWCOUPLER(R), 3.5 mm, STERILE EO, Rx only, 1 FLOWCOUPLER, REF GEM2755-FC…Instructions for use booklet may puncture the outer Tyvek lid. 04/21/2016Terminated
Z-1631-2016 Class IIGEM(TM), FLOWCOUPLER(R), 4.0 mm, STERILE EO, Rx only, 1 FLOWCOUPLER, REF GEM2756-FC…Instructions for use booklet may puncture the outer Tyvek lid. 04/21/2016Terminated