Recall Z-1629-2015

Siemens Healthcare Diagnostics, Inc. · initiated April 9, 2015

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

IMMULITE/IMMULITE 1000 Systems TBG, catalog LKTB1 Quantitative measurement of Thyroxine-Binding Globulin (TBG) in serum, as an aid in the diagnosis and treatment of thyroid diseases.

Codes / lots: LOT 0257, SMN 10381643

Reason for recall

IMMULITE/IMMULITE 1000 TBG catalog LKTB lot 257 is failing the adjustment slope instrument parameter of <1.80. The resulting slope of the kit is >1.80 resulting in the potential to produce depressed TBG patient values.

FDA-determined cause: Unknown/Undetermined by firm

Action taken

Siemens sent an Urgent Field Safety Notice dated April 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Actions to be Taken by the Customer Discontinue use and discard kit lot listed in Table 1. Review your inventory of these products to determine your laboratory's replacement needs. Please review this letter with your Medical Director. Complete and return the Field Correction Effectiveness Check Form attached to this letter within thirty (30) days. Please retain this letter with your laboratory records, and forward this letter to those who may have received this product. We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Customer Care Center or your local Siemens technical support representative. For further questions call (914) 631-8000.

Details

Recall number
Z-1629-2015
Initiated
April 9, 2015
Posted by FDA
May 19, 2015
Terminated
March 22, 2017
Product code
CEE – Radioimmunoassay, Thyroxine-Binding Globulin
Quantity
155 kits
Distribution
Worldwide Distribution to the countries of : Denmark, Spain, Great Britain, Germany, Bosnia Herzeg, China and Russian Federation.
Recalling firm
Siemens Healthcare Diagnostics, Inc., Tarrytown, NY

Official FDA recall record