Recall Z-1629-2018

Abbott Gmbh & Co. KG · initiated December 14, 2017

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

ARCHITECT DHEA-S Reagent Product Usage: It is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of dehydroepiandrosterone sulfate (DHEA-S) in human serum and plasma on the ARCHITECT iSystem.

Codes / lots: List No. 8K2-20 with the following Lot No., Exp. Date, UDI: a. Lot No. 01316K000, Exp. Date 04FEB2018, UDI (01) 00380740017446 (17) 180204 (10) 01316K000 (240) 8K2720; b. Lot No. 01217A000, Exp. Date 13APR2018, UDI (01) 00380740017446 (17) 180413 (10) 01217A000 (240) 8K2720; c. Lot No. 02217D000, Exp. Date 24JUN2018, UDI (01) 00380740017446 (17) 180624 (10) 02217D000 (240) 8K2720; d. Lot No. 04117H000, Exp. Date 25NOV2018, UDI (01) 00380740017446 (17) 181125 (10) 04117H000 (240) 8K2720; List No. 8K27-25 with the following Lot No., Exp. Date, UDI: a. Lot No. 01216K000, Exp. Date 04FEB2018, UDI (01) 00380740017453 (17) 180204 (10) 01216K000 (240) 8K2725; b. Lot No. 01117A000, Exp. Date 13APR2018, UDI (01) 00380740017453 (17) 180413 (10) 01117A000 (240) 8K2725; c. Lot No. 02217C000, Exp. Date 17JUN2018, UDI (01) 00380740017453 (17) 180617 (10) 02217C000 (240) 8K2725; d. Lot No. 02317D000, Exp. Date 24JUN2018, UDI (01) 00380740017453 (17) 180624 (10) 02317D000 (240) 8K2725; e. Lot No. 02417D000, Exp. Date 24JUN2018, UDI (01) 00380740017453 (17) 180624 (10) 02417D000 (240) 8K2725; f. Lot No. 00517F000, Exp. Date 02SEP2018, UDI (01) 00380740017453 (17) 180902 (10) 00517F000 (240)…

Reason for recall

Falsely elevated results may be obtained when using the ARCHITECT DHEA-S assay with samples from infants up to 60 days old.

FDA-determined cause: Device Design

Action taken

Customers were notified via letter on about 12/14/2017. Instructions included to immediately discontinue use of the ARCHITECH DHEA-S assay with samples form infants up to 60 days old, to immediately discontinue use of the ARCHITECH DHEA-S reagent insert Expected Values representative data for children up to 10 years old, to review the letter with the Medial Director and follow the laboratory protocol regarding the need for reviewing previously reported patient results, to forward the Product Correction letter to customers if the product was further distributed, and to complete and return the Customer Reply Form. Questions from U.S. customers can be directed to Customer Service at 1-877-4ABBOTT (available 24 hours per day, 7 days a week) and customers outside the U.S. should contact their local area Customer Service.

Details

Recall number
Z-1629-2018
Initiated
December 14, 2017
Terminated
October 10, 2018
Product code
JKC – Radioimmunoassay, Dehydroepiandrosterone (Free And Sulfate)
Quantity
10,153 units
Distribution
Worldwide Distribution in the states of AL, AZ, CA, CO, CT, FL, GA, IL, IN, KS, KY, LA, MA, MN, MO, MS, NC, NJ, NY, OH, OK, PA, PR, SC, SD, TN, TX, UT, VT, WA, WY. Distributed internationally to Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Barbados, Belarus, Belgium, Bosnia and Herzegovi, Botswana, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica…
Recalling firm
Abbott Gmbh & Co. KG, Wiesbaden

Official FDA recall record