Recall Z-1634-2020

Smith & Nephew, Inc. · initiated February 27, 2020

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

NAVIO Soft Tissue Protector, included in instrument kits. The NAVIO Tissue Protector is a reusable instrument designed to, when used as instructed, provide the required bone pin spacing for the Navio bone clamps and potentially reduce the amount of patient tissue affected when placing the bone pins.

Codes / lots: Component Product Number: PFSR101092; a. Instrument Kit Number: PFSR02050, Lot Numbers 2067, 2212, 3809, 3869, 4354, 5228, 5337, 5494, 5514, 5887, 6248; b. Instrument Kit Number: PFSR02051, Lot Numbers 4354, 5514; c. Instrument Kit Number: PFSR02052, Lot Numbers 4354, 5494, 6248; d. Instrument Kit Number: PFSR02053, Lot Numbers 6248

Reason for recall

Potential for the NAVIO Soft Tissue Protector to become stuck or bound to the bone pin intraoperatively when inserting the NAVIO bone pins.

FDA-determined cause: Device Design

Action taken

The firm, Smith & Nephew, notified customers on about 03/03/2020 via "Urgent Medical Device Recall Notice" letter. The customers were Instructed to examine inventory and quarantine affected devices, notify customers if further distributed, and complete and return the Response Form to fieldactions@smith-nephew.com or fax to 1-901-566-7975. Questions and concerns can be directed to EieldActions@smlth-nephew .com. or 901-399-1970.

Details

Recall number
Z-1634-2020
Initiated
February 27, 2020
Terminated
August 2, 2024
Product code
OLO – Orthopedic Stereotaxic Instrument
Quantity
348 units
Distribution
Nationwide distribution to AL, AZ, CA, CT, FL, GA, HI, IA, IL, KS, LA, MA, MI, MN, MS, MT, NC, NJ, NY, OH, OK, PA, TN, TX, VA, WA, WI, WV. International distribution to Turkey, United Kingdom, Thailand, Switzerland, France, Australia.
Recalling firm
Smith & Nephew, Inc., Memphis, TN