Recall Z-1649-2021

Biopsybell S.R.L. · initiated March 22, 2021

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

10 GAUGE DRILL, CODE INTVM-DRL

Codes / lots: LOT: 20102486

Reason for recall

Products labeled as sterile were distributed, but may not have been sterilized.

FDA-determined cause: Vendor change control

Action taken

The firm notified two consignees with the preliminary information by email on 03/22/2021 and followed with the official notice with the complete information on 04/08/2021 to all consignees. The letter requested that the consignee cease using the affected product and contact Biopsybell or a Biopsybell representative for product return/disposal and related replacement. Distributors were directed to identify those who received the product and notify them.

Details

Recall number
Z-1649-2021
Initiated
March 22, 2021
Posted by FDA
May 19, 2021
Terminated
February 17, 2026
Product code
LOD – Bone Cement
Quantity
75 units
Distribution
US Nationwide distribution in the states MS and KY.
Recalling firm
Biopsybell S.R.L., Mirandola

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1631-2021 Class IIBPB medica CEMIX, PRODUCT NUMBER KVTMIXN, Vertebroplasty-Kyphoplasty cement mixing systemProducts labeled as sterile were distributed, but may not have been sterilized. 03/22/2021Terminated
Z-1632-2021 Class IIRENOVA SPINE BONE FILLER 10G 12CM CENTIM. STERILE, RESFIL1112C1-USProducts labeled as sterile were distributed, but may not have been sterilized. 03/22/2021Terminated
Z-1633-2021 Class IIBPB medica RENOVA SPINE KYPHOPLASTY WORKING CANNULA AND TROCAR TIP, 187 mm, 10G, RES…Products labeled as sterile were distributed, but may not have been sterilized. 03/22/2021Terminated
Z-1634-2021 Class IIBPB medica OSTEOBELL EXPLANT NEEDLE FOR BONE MARROW EXPLANT, 110 mm, 11 G, REF OBE1111Products labeled as sterile were distributed, but may not have been sterilized. 03/22/2021Terminated
Z-1635-2021 Class IICEMENT MIXING SYSTEM, VERTEBROPLASTY-KYPHOPLASTY, REF VCF-1051Products labeled as sterile were distributed, but may not have been sterilized. 03/22/2021Terminated
Z-1636-2021 Class IIEASYNJECT, Device for cement infusion in the vertebroplasty procedure, REF KVTGUN-DSProducts labeled as sterile were distributed, but may not have been sterilized. 03/22/2021Terminated
Z-1637-2021 Class II10 GAUGE DIRECT UNILATERAL STYLET(1 X DIAMOND 1 X BEVEL), CODE INTVM-DSSKBDProducts labeled as sterile were distributed, but may not have been sterilized. 03/22/2021Terminated
Z-1638-2021 Class II10 GAUGE CEMENT DELIVERY CANNULA, CODE INTVM-CDCProducts labeled as sterile were distributed, but may not have been sterilized. 03/22/2021Terminated
Z-1639-2021 Class II10G DIRECT SINGLE - ACCESS, CODE INTVM-DSBAKProducts labeled as sterile were distributed, but may not have been sterilized. 03/22/2021Terminated
Z-1640-2021 Class II8G DIRECT WORKING CANNULA BEVEL TIP, CODE INTVM-DWCBProducts labeled as sterile were distributed, but may not have been sterilized. 03/22/2021Terminated
Z-1641-2021 Class II11G DIRECT SINGLE - DELIVERY (4 FILLERS-4 SYRINGES), CODE INTVMN-SCDKProducts labeled as sterile were distributed, but may not have been sterilized. 03/22/2021Terminated
Z-1642-2021 Class IICEMIX - CEMENT MIXING SYSTEM, CODE INTV-MMSProducts labeled as sterile were distributed, but may not have been sterilized. 03/22/2021Terminated
Z-1643-2021 Class II11G DIRECT WORKING CANNULA DIAMOND TIP MINI, CODE INTVMN-DWCDProducts labeled as sterile were distributed, but may not have been sterilized. 03/22/2021Terminated
Z-1644-2021 Class II11G DIRECT SINGLE - ACCESS, CODE INTVMN-DSBAKProducts labeled as sterile were distributed, but may not have been sterilized. 03/22/2021Terminated
Z-1645-2021 Class II8G DIRECT WORKING CANNULA BEVEL TIP, CODE INTV-DWCBProducts labeled as sterile were distributed, but may not have been sterilized. 03/22/2021Terminated
Z-1646-2021 Class II10 GAUGE SINGLE CEMENT DELIVERY 4 FILLERS 4 SYRINGES, CODE INTVM-SCDKProducts labeled as sterile were distributed, but may not have been sterilized. 03/22/2021Terminated
Z-1647-2021 Class II10G DIRECT WORKING CANNULA BEVEL TIP 15CM, CODE INTVM-DWCB/15Products labeled as sterile were distributed, but may not have been sterilized. 03/22/2021Terminated
Z-1648-2021 Class II10G COAXIAL CANNULA WITH TUOHY CONNECTION, CODE INTVM-CCProducts labeled as sterile were distributed, but may not have been sterilized. 03/22/2021Terminated