Recall Z-1657-2026

Siemens Healthcare Diagnostics, Inc. · initiated February 23, 2026

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Atellica CH Enzymatic Hemoglobin A1c (A1c_E); Siemens Material Number (SMN): 11097536;

Codes / lots: Siemens Material Number (SMN): 11097536; UDI-DI: 00630414220505; Lot numbers: 150131, 150132, 150232, 150260, 150283, 150344, 150361, 150365, 150366, 150445, 150453, 150460, 150498, 150505, 150562, 150600, 150618, 150621;

Reason for recall

A potential for depressed results for Atellica CH A1c_E when processing Atellica CH RCRP on the same analyzer. Customers processing A1c_E and RCRP on separate analyzers are not impacted by the recall. The frequency and magnitude of the negative bias and imprecision increase over time and are not consistent with the assay Instructions For Use (IFU). When this issue is present, Quality Control (QC) results demonstrate a negative trend and increased imprecision. The clinical impact of a negative (A1c_E) bias may lead to a delay in diagnosis for diabetes or lead to less-than-optimal treatment for existing diabetes.

FDA-determined cause: Process control

Action taken

On February 23, 2026 Urgent Medical Device Correction letters were sent to customers. Actions required: If your laboratory is not processing RCRP and A1c_E on the same analyzer, no action is required. If your laboratory is processing RCRP and A1c_E on the same analyzer, monitor A1c_E QC for evidence of negative bias and increased imprecision. Please review this letter with your Medical Director to determine the appropriate course of action, including evaluation of any previously generated results, if applicable. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. Retain this letter with your laboratory records and forward it to others who may have received this product. Resolution A follow-up communication will be provided when Customer Actions are no longer required. We appreciate your continued partnership and apologize for any inconvenience this situation may cause. If you have any questions or require assistance, please contact your Siemens Healthineers Customer Care Center or your local technical support representative.

Details

Recall number
Z-1657-2026
Initiated
February 23, 2026
Posted by FDA
March 27, 2026
Product code
PDJ – Hemoglobin A1c Test System
Quantity
21,291 units (9750 US, 11541 OUS)
Distribution
Domestic: Nationwide Distribution International: Argentina, Australia, Bangladesh, Brazil, Bulgaria , Canada, Chile, China, Colombia, Croatia , Cura¿ao, St Eus, Czech Republic , Denmark , Ecuador, Egypt , Estonia , Finland , France , Georgia , Germany , Greece , Hong Kong, India, Iraq , Italy , Jordan , Kuwait , Latvia , Lithuania , Macedonia , Malaysia, Maldives, Mexico, Netherlands , New…
Recalling firm
Siemens Healthcare Diagnostics, Inc., Tarrytown, NY

Official FDA recall record