Recall Z-1669-2014

Synthes, Inc. · initiated April 29, 2014

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Synthes ProDisc-C 2.0mm Milling Bit, Synthes 2.0mm Milling Bit HXC, Synthes 2.0mm Milling Bit with 2.35mm x 33.5mm Shaft, and Synthes 2.0mm Milling Bit with Flat Coupling. Spinal surgical instrument: The Synthes ProDisc-C 2.0mm Milling Bits are used as a part of the milling option for keel preparation of the vertebral bodies. The milling bits are coupled to a power tool and interface with the milling guide, which is connected to a trial implant for the keel preparation step.

Codes / lots: part number: 03.820.161s, with lot numbers: 5742507,5755970,5755971,5755972, 5755973,5755974,5828298,6336115, 6338099,6338100,6340029,6340030, 6344199,6344200,6363601,6385892, 6407046,6431260,6431261,6460402, 6503360,6526127,6571070,6571073, 6571074,6601985,6636946,6636947, 6636948,6653819,6653820,6670383, 6691341,6691342,6718933,6718934, 6718935,6740611,6794651,6794652, 6794653,6794654,6794655 part number: 03.820.1638 with lot numbers: 5742506,5755978,5755979,5755980, 5755981,5755982,5828299,5828300, 6320586,6336108,6336110,6336111, 6340033,6344201,6344202,6363602, 6385893,6431262,6431263,6460403, 6503361,6571075,6601986,6601987, 6636949,6636950,6636951,6653810, 6653818,6670382,6691343,6691344, 6718938,6740615,6740616,6745821, 6770474,6770475,6770476,6770477, 6770478,6794657,6794658 part number: 03.820.167S with lot numbers: 5755983,5755984,5755987,5755988, 5755989,5755985,5755986,5755990, 5828302,6338101,6338102,6338103, 6338104,6338106,6338107,6338108, 6338109,6344205,6471089,6670381, 6718939,6718940,6718941,6843907, 6843906,6856252,6856253,6867662, 6867661,6901712,6901713,6919726, 6919727,6919728,6901710,6901711, 6921717,6921718,6940357,6960192…

Reason for recall

Certain lots of the ProDisc-C 2.0mm Milling Bits have an increased potential to break during use.

FDA-determined cause: Process control

Action taken

DePuy Synthes sent a Recall letter, dated April 29, 2014, to affected customers. The letter identified the affected customers, problem and actions to be taken. The letter instructed customers to discontinue use and distribution of the device and to return affected devices to Synthes. For questions call 610-719-5450.

Details

Recall number
Z-1669-2014
Initiated
April 29, 2014
Posted by FDA
May 29, 2014
Terminated
May 19, 2016
Product code
GFG – Bit, Surgical
Quantity
6690
Distribution
US Nationwide Distribution
Recalling firm
Synthes, Inc., West Chester, PA

Official FDA recall record