Recall Z-1673-2025
Siemens Healthcare Diagnostics Inc · initiated March 31, 2025
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Stratus CS Acute Care cTNI TestPak.IVD test for cardiac troponin I in heparinized plasma.
Codes / lots: Siemens Materials No. 10445071; UDI-DI: 0405686902174VK; Lot No. 234337002 and forward.
Reason for recall
Siemens Healthcare Diagnostics Inc., is recalling their Stratus CS Acute Care cTNI TestPak, an IVD test for cardiac troponin I in heparinized plasma by removal. The reason for the recall is an increased occurrence of random non-repeatable false positive cardiac Troponin l(cTnI) results at any point during the TestPaks shelf life when using the Stratus CS cTnI Acute Care Testpak.
FDA-determined cause: Under Investigation by firm
Action taken
An Urgent Medical Device Correction notice dated March 2025 was mailed to consignees. The notice instructed consignees to review the provided notification with their medical director to determine the appropriate course of action. Siemens will reimburse users for repeat tests and discarded units associated with this recall; consignees should contact the Customer Care Center for reimbursement. If product was further distributed, Siemens asks that the recall notification be forwarded. The recall notification is to be retained with laboratory records and forwarded to those who use the recalled devices. Consignees with any questions can contact Siemens Healthineers Technical Solutions Center at 977-229-3711.
Details
- Recall number
- Z-1673-2025
- Initiated
- March 31, 2025
- Posted by FDA
- April 25, 2025
- Product code
- JHX – Fluorometric Method, Cpk Or Isoenzymes
- Quantity
- 367 units
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, IL, KS, MI, MO, NJ, NY, OH, OK, PA, SC & WI. The countries of Austria, Canada, Chile, France, Germany, Iraq, Italy, Malaysia, North Macedonia, Saudi Arabia, Serbia, Slovenia, Spain, Sweden, Switzerland & Vietnam.
- Recalling firm
- Siemens Healthcare Diagnostics Inc, Norwood, MA