Recall Z-1674-2009

Essential Medical Supply, Inc. · initiated May 8, 2009

Terminated

Product

Essential Medical Supply brand Essential Rollator with Loop Hand Brakes, models W1700B (Blue) and W1700R (Red), Four Wheel Aluminum Walker/Rollator

Codes / lots: W1700 (R and B); Lot / Serial Nos.: KLD-E07090001 through KLD-E08020570

Reason for recall

The caster may separate from the frame completely causing injury to the user.

FDA-determined cause: Device Design

Action taken

Essential Medical Supply Inc issued a Notice of Quarantine Letter to all customers on 04/30/09, identifying affected products as W1700B & W1700R Essential Rollators, Serial #'s KLD-E07090001 through KLD-E08020570, received after 11/05/2007. Firm issued a Device Recall Letter to all direct customers on 05/08/09 instructing them to do the following: (1) segregate and quarantine any affected product; (2) identify and notify any customers that may received the affected products, instructing them to return the product to their dealer; and (3) either return the affected products to Essential Medical Supply or destroy the products and send proof of destruction.

Details

Recall number
Z-1674-2009
Initiated
May 8, 2009
Posted by FDA
July 20, 2009
Terminated
September 15, 2010
Product code
ITJ – Walker, Mechanical
Quantity
2265
Distribution
Nationwide distribution.
Recalling firm
Essential Medical Supply, Inc., Orlando, FL