Recall Z-1679-2026

Philips Medical Systems Nederland B.V. · initiated March 3, 2026

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Azurion 5 M12; System Code: (1)722227, (2)722231;

Codes / lots: (1) System Code: 722227; UDI: 00884838099227; System Serial Number: 265, 135, 88, 276, 264, 271, 63, 238, 360, 23, 127, 159, 197, 149, 54, 56, 230, 192, 156, 250, 98, 31, 128, 81, 24, 37, 320, 179, 270, 151, 53, 246, 245, 84, 38, 28, 29, 170, 30, 338, 324, 108, 305, 168, 172, 356, 346, 62, 236, 25, 313, 339, 104, 321, 59, 109, 101, 19, 155, 99, 285, 247, 261, 298, 177, 138, 337, 157, 158, 228, 249, 61, 287, 55, 239, 306, 215, 336, 213, 76, 216, 73, 318, 353, 65, 167, 41, 85, 191, 20, 226, 102, 40, 87, 163, 83, 251, 268, 309, 310, 316, 290, 272, 5, 6, 7, 218, 36, 75, 195, 43, 345, 116, 18, 153, 351, 352, 14, 15, 120, 262, 204, 258, 326, 348, 17, 141, 114, 103, 355, 113, 110, 315, 21, 161, 194, 105, 173, 174, 64, 49, 71, 91, 187, 166, 210, 219, 222, 273, 205, 280, 300, 256, 335, 121, 80, 123, 243, 344, 308, 22, 140, 90, 142, 69, 200, 72, 362, 242, 323, 331, 136, 269, 137, 77, 89, 294, 181, 234, 95, 2, 111, 221, 293, 165, 342, 1, 12, 16, 257, 330, 67, 340, 260, 148, 281, 284, 248, 333, 199, 183, 106, 253, 145, 164, 241, 52, 26, 233, 301, 211, 154, 217, 332, 334, 44, 198, 297, 206, 292, 130, 58, 203, 47, 48, 107, 93, 125, 240, 122, 118, 112, 46, 176, 92, 252, 184, 57, 202, 100, 115…

Reason for recall

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

FDA-determined cause: Device Design

Action taken

On March 3, 2026, URGENT Medical Device Correction letters were sent to customers. Actions to be taken: 1. Circulate this Urgent Medical Device Correction letter to all users of the system so that they are aware of the issue and follow the instructions below. 2. Perform the daily check before starting the first procedure, as described in the device labelling 3. Should you experience a foot switch activation issue, please report it to your local Philips representative. 4. In case the affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction letter to that organization and inform Philips about this transfer through your local Philips representative. 5. Keep this Urgent Medical Device Correction letter with the documentation of the system until Philips corrects your system. Ensure that the letter is in a place likely to be seen/viewed. 6. Complete and return the response form included in this Urgent Medical Device Correction letter to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction letter and understanding of the issue and required actions to be taken.

Details

Recall number
Z-1679-2026
Initiated
March 3, 2026
Posted by FDA
March 27, 2026
Product code
OWB – Interventional Fluoroscopic X-Ray System
Quantity
345 units (10 US, 335 OUS)
Distribution
Domestic: Nationwide Distribution; International: Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican…
Recalling firm
Philips Medical Systems Nederland B.V., Best

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1659-2026 Class IIAllura Xper FD10C; System Code: 722001;Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when… 03/03/2026Open, Classified
Z-1660-2026 Class IIALLURA Xper FD10F; System Code: 722002;Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when… 03/03/2026Open, Classified
Z-1661-2026 Class IIALLURA Xper FD10; System Code: (1) 722003, (2)722010, (3)722026;Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when… 03/03/2026Open, Classified
Z-1662-2026 Class IIALLURA Xper FD10/10; System Code: (1) 722005, (2)722011, (3)722027;Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when… 03/03/2026Open, Classified
Z-1663-2026 Class IIALLURA Xper FD20; System Code: (1) 722006, (2)722012, (3)722028;Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when… 03/03/2026Open, Classified
Z-1664-2026 Class IIALLURA Xper FD20 Biplane; System Code: (1) 722008, (2)722013;Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when… 03/03/2026Open, Classified
Z-1665-2026 Class IIALLURA Xper FD20 OR Table; System Code: (1) 722015, (2)722023, (3)722035;Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when… 03/03/2026Open, Classified
Z-1666-2026 Class IIALLURA Xper FD20 Biplane OR Table; System Code: (1) 722020, (2)722025;Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when… 03/03/2026Open, Classified
Z-1667-2026 Class IIALLURA Xper FD10 OR Table; System Code: (1) 722022, (2)722033;Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when… 03/03/2026Open, Classified
Z-1668-2026 Class IIAllura Xper FD20/10; System Code: 722029;Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when… 03/03/2026Open, Classified
Z-1669-2026 Class IIAllura Xper FD20/20; System Code: 722038;Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when… 03/03/2026Open, Classified
Z-1670-2026 Class IIAllura Xper FD20/20 OR Table; System Code: 722039;Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when… 03/03/2026Open, Classified
Z-1671-2026 Class IIAllura Xper FD20/15; System Code: 722058;Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when… 03/03/2026Open, Classified
Z-1672-2026 Class IIAllura Xper FD20/15 OR Table; System Code: 722059;Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when… 03/03/2026Open, Classified
Z-1673-2026 Class IIAzurion 3 M12; System Code: (1)722063, (2)722221, (3)722229;Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when… 03/03/2026Open, Classified
Z-1674-2026 Class IIAzurion 3 M15; System Code: (1)722064, (2)722222, (3)722230, (4)722280 (OUS Only);Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when… 03/03/2026Open, Classified
Z-1675-2026 Class IIAzurion 7 B12; System Code: (1)722067, (2)722225, (3)722235;Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when… 03/03/2026Open, Classified
Z-1676-2026 Class IIAzurion 7 B20; System Code: (1)722068, (2)722226, (3)722236;Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when… 03/03/2026Open, Classified
Z-1677-2026 Class IIAzurion 7 M12; System Code: (1)722078, (2)722223, (3)722233;Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when… 03/03/2026Open, Classified
Z-1678-2026 Class IIAzurion 7 M20; System Code: (1)722079, (2)722224, (3)722234, (4)722282 (OUS only);Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when… 03/03/2026Open, Classified
Z-1680-2026 Class IIAzurion 5 M20; System Code: (1)722228, (2)722232, (3) 722281 (OUS only);Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when… 03/03/2026Open, Classified