Product

InSound Medical, Inc, Lyric in-canal hearing aid.

Codes / lots: All devices manufactured between 12/10/2009 and 12/24/2009. Expiration dates 12/10/2010 through 12/24/2010.

Reason for recall

Manufacturing error could result in electrolyte leakage from the product's battery.

FDA-determined cause: Process design

Action taken

InSound sent to its consignees a "Dear Lyric Provider" letter of explanation on 2/1/2010.

Details

Recall number
Z-1681-2010
Initiated
January 19, 2010
Posted by FDA
May 26, 2010
Terminated
August 2, 2011
Product code
ESD – Hearing Aid, Air-Conduction, Prescription
Quantity
2019 units
Distribution
Distribution only in US ( nationwide). No foreign consignees involved.
Recalling firm
InSound Medical, Inc., Newark, CA