Recall Z-1684-2017

Streck · initiated February 23, 2017

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Cyto-Chex BCT - blood specimen collection device - Immunophenotyping Preservative. IVD. Sterile. 2.0ml only. Packaged 6-tube pack REF 213559 and 100-tube box REF 218980. Streck.

Codes / lots: UDI Number 10844509000836 and 20844509000833. Lot(s) Expiration Date 50910309 04-01-2017 60760309 03-16-2018 61460309 05-25-2018

Reason for recall

Complaints of the 2.0ml tube stopper coming loose during phlebotomy when small diameter tube adapters were not utilized. Such adapters aid with centering small diameter tubes to properly pierce the middle of the stoppers.

FDA-determined cause: Error in labeling

Action taken

On February 23, 2017, Streck electronically notified customers of the updated device correction. The recalling firm used an electronic mail alert to notify all of their customers the IFU for the recalled product had been revised. The specified the revision included phlebotomy guidelines specific to the 2.0 ml blood collections tubes. The revised IFU was provided as a link in the email. The recalling firm is reportedly able to monitor the access of the revised IFU to ensure each customer has received the information. For further questions, please call (402) 333-1982.

Details

Recall number
Z-1684-2017
Initiated
February 23, 2017
Posted by FDA
March 8, 2017
Product code
GIM – Tubes, Vacuum Sample, With Anticoagulant
Quantity
128,692 tubes
Distribution
Worldwide Distribution - USA (nationwide) and to the countries of : Belgium, Canada, Germany, Israel, Japan, the Republic of Korea, the Netherlands, and the United Kingdom. There was no military or government distribution.
Recalling firm
Streck, La Vista, NE

Official FDA recall record