Recall Z-1686-2022

Datascope Corporation · initiated August 5, 2022

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Sensation Plus 8Fr 50cc Intra-Aortic Balloon Catheter

Codes / lots: Code information reads "UDI-DI, Lot No., Finished Goods Part No.": 10607567108605, 3000168086, 0684-00-0576-01; 10607567108605, 3000217952, 0684-00-0576-01; 10607567109633, 3000166607, 0684-00-0576-01U; 10607567109633, 3000169722, 0684-00-0576-01U; 10607567109633, 3000172352, 0684-00-0576-01U; 10607567109633, 3000172353, 0684-00-0576-01U; 10607567109633, 3000176574, 0684-00-0576-01U; 10607567109633, 3000176633, 0684-00-0576-01U; 10607567109633, 3000185680, 0684-00-0576-01U; 10607567109633, 3000196023, 0684-00-0576-01U; 10607567109633, 3000199196, 0684-00-0576-01U; 10607567109633, 3000199197, 0684-00-0576-01U; 10607567109633, 3000199198, 0684-00-0576-01U; 10607567109633, 3000200147, 0684-00-0576-01U; 10607567109633, 3000213850, 0684-00-0576-01U

Reason for recall

Certain lots containing undersized dilator.

FDA-determined cause: Vendor change control

Action taken

The firm distributed an Urgent Medical Device Removal notification, by mail, dated August 5, 2022. The notification informs customers that certain lots of 8Fr. 50cc IAB catheter and insertion kits are being recalled due to an undersized vessel dilator. During a sheathless insertion, the undersized inner diameter may result in the inability to insert the guidewire through the inner lumen of the dilator and the undersized outer diameter may reduce dilation of the vessel. If used in a procedure, recalled product could result in a procedural delay, gas emoli, and vessel injury. Customers are asked to examine their inventory to determine if they have nay recalled product and to quarantine any affected units. Customers with recalled product are to contact Datascope Corp./Getinge Customer Service at 888-9GETUSA (888-943-8872, option 2) between the hours of 8 AM and 6 PM to request a return authorization number to receive full credit. Distributors who have shipped any product are to forward the notification to customers. The recall notification is to be forwarded to all users at hospitals/facilities. Customers are asked to complete and return the recall response form by email to Dilator8Fr2022@getinge.com or by fax to 877-647-1921.

Details

Recall number
Z-1686-2022
Initiated
August 5, 2022
Posted by FDA
September 8, 2022
Product code
DSP – System, Balloon, Intra-Aortic And Control
Quantity
10,427 units
Distribution
Awaiting domestic distribution. Foreign: Australia, Belgium, Canada, Czech Republic, France, Germany, Hong Kong, Iceland, Ireland, Italy, Netherlands, New Zealand, Norway, Poland, Portugal, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.
Recalling firm
Datascope Corporation, Fairfield, NJ

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1685-2022 Class IIMEGA 8Fr 50cc Intra-Aortic Balloon CatheterCertain lots containing undersized dilator. 08/05/2022Open, Classified
Z-1687-2022 Class IIInsertion Kit for use with SENSATION PLUS 8F. 50cc Intra-Aortic Balloon CatheterCertain lots containing undersized dilator. 08/05/2022Open, Classified