Recall Z-1692-2022

American Contract Systems, Inc. · initiated June 16, 2022

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Various PICC Line, Central Line Packs

Codes / lots: Tray Number, Sterilization Lot, Lot Number, Bag Serial Number 1) LSRPICCLINEH, 2105051, 884211, 14886484 2) UICL44M, 2107142, 807211, 64164800

Reason for recall

Product was exposed to multiple sterilization cycles without validation for multiple exposures.

FDA-determined cause: Process change control

Action taken

The recalling firm began notifying customers of the recall beginning June 13, 2022. Customer notification was initially made via email and an "Urgent Medical Device Recall" letter dated June 16, 2022. Customers are asked to review inventory and segregate affected product, which should be identified with the Bag Serial Number. Affected product should be destroyed, with evidence of destruction provided back to the recalling firm. A Field Action Response Form should be completed and returned to angela.bunn@owens-minor.com as soon as possible, even if the customer has no inventory. If you have any questions, please call 816-920-5846.

Details

Recall number
Z-1692-2022
Initiated
June 16, 2022
Product code
OFF – Central Venous Catheter Tray
Quantity
2 units
Distribution
Distribution within the United States to health systems and distributors in AZ, FL, IA, IL, MA, MN, MO, OH, NE, PA, RI, SD, and TX
Recalling firm
American Contract Systems, Inc., Kansas City, MO

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1689-2022 Class IIVarious Angiography packs/traysProduct was exposed to multiple sterilization cycles without validation for multiple exposures. 06/16/2022Open, Classified
Z-1690-2022 Class IIVarious Angiography/Cath Lab packs/traysProduct was exposed to multiple sterilization cycles without validation for multiple exposures. 06/16/2022Open, Classified
Z-1691-2022 Class IIVarious Pacemaker Packs, Cardio Thoracic PacksProduct was exposed to multiple sterilization cycles without validation for multiple exposures. 06/16/2022Open, Classified
Z-1693-2022 Class IIVarious ENT PacksProduct was exposed to multiple sterilization cycles without validation for multiple exposures. 06/16/2022Open, Classified
Z-1694-2022 Class IIVarious Laparoscopy PacksProduct was exposed to multiple sterilization cycles without validation for multiple exposures. 06/16/2022Open, Classified
Z-1695-2022 Class IIVarious Cysto PacksProduct was exposed to multiple sterilization cycles without validation for multiple exposures. 06/16/2022Open, Classified
Z-1696-2022 Class IIVarious Port Insertion PacksProduct was exposed to multiple sterilization cycles without validation for multiple exposures. 06/16/2022Open, Classified
Z-1697-2022 Class IIBreast Abdominoplasty PackProduct was exposed to multiple sterilization cycles without validation for multiple exposures. 06/16/2022Open, Classified
Z-1698-2022 Class IIVarious convenience packsProduct was exposed to multiple sterilization cycles without validation for multiple exposures. 06/16/2022Open, Classified
Z-1699-2022 Class IIVarious convenience packsProduct was exposed to multiple sterilization cycles without validation for multiple exposures. 06/16/2022Open, Classified
Z-1700-2022 Class IIVarious neurology packsProduct was exposed to multiple sterilization cycles without validation for multiple exposures. 06/16/2022Open, Classified
Z-1701-2022 Class IIVarious ob/gyn packsProduct was exposed to multiple sterilization cycles without validation for multiple exposures. 06/16/2022Open, Classified
Z-1702-2022 Class IIVarious hysterectomy, vaginal packsProduct was exposed to multiple sterilization cycles without validation for multiple exposures. 06/16/2022Open, Classified
Z-1703-2022 Class IIVarious C-section packsProduct was exposed to multiple sterilization cycles without validation for multiple exposures. 06/16/2022Open, Classified
Z-1704-2022 Class IID&C packProduct was exposed to multiple sterilization cycles without validation for multiple exposures. 06/16/2022Open, Classified
Z-1705-2022 Class IIVarious eye packs, cataract packsProduct was exposed to multiple sterilization cycles without validation for multiple exposures. 06/16/2022Open, Classified