Recall Z-1699-2023
C.R. Bard Inc · initiated April 20, 2023
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Magic3 Go Intermittent Urinary Catheters, male Catalog Numbers: a) REF 50810G, b) REF 50812G, c) REF 50814G, d) REF 50816G, e) REF 50818G, f) REF 53810G, g) REF 53812G, h) REF 53814G, i) REF 53816G, j) REF 53818G, k) REF 53820G
Codes / lots: a) REF 50810G, UDI/DI 0 801741137006; Lot Numbers: JUGV1658; b) REF 50812G, UDI/DI 0 801741137013; Lot Numbers: JUGU0844, JUGV1290, JUGW9017; c) REF 50814G, UDI/DI 0 801741137020; Lot Numbers: JUGS1485, JUGT0238, JUGT0980, JUGT1000, JUGT1710, JUGT2735, JUGT3299, JUGT3353, JUGU0874, JUGU1917, JUGU1956, JUGV0295, JUGV0665, JUGV1277, JUGV1682, JUGW0073, JUGW0812, JUGW1444, JUGW1859; d) REF 50816G, UDI/DI 0 801741137037; Lot Numbers: JUGT0239, JUGT0289, JUGT1010, JUGT1690, JUGT2703, JUGT3399, JUGU0783, JUGU1969, JUGU9049, JUGV0317, JUGV0666, JUGV0690, JUGV9055, JUGX9006; e) REF 50818G, UDI/DI 0 801741137044; Lot Numbers: JUGT9170, JUGT9173, JUGU1788; f) REF 53810G, UDI/DI 0 801741136948; Lot Numbers: JUGQ0944, JUGQ3276, JUGT2737, JUGT9167, JUGU9029, JUGU9035, JUGU9038, JUGU9044, JUGU9051, JUGV9053, JUGX1267, JUGX1988; g) REF 53812G, UDI/DI 0 801741136955; Lot Numbers: JUGQ0939, JUGQ3264, JUGR1444, JUGR2031, JUGS0633, JUGS2403, JUGS3215, JUGT1012, JUGT1711, JUGT2705, JUGT3400, JUGU0845, JUGU1919, JUGV0667, JUGV1292, JUGV9148, JUGW0794; h) REF 53814G, UDI/DI 0 801741136962; Lot Numbers: JUGQ3265, JUGR0224, JUGR1388, JUGR2051, JUGS0048, JUGS0610, JUGS1399, JUGS2381, JUGS3219, JUGT0299…
Reason for recall
Some catheters had a manufacturing defect where the clear resealing label separating from the product foil pouch, resulting in a possible opening into the product packaging that could impact product sterility.
FDA-determined cause: Under Investigation by firm
Action taken
The firm issued an URGENT MEDICAL DEVICE RECALL to its consignee beginning on 04/20/2023 by Fed Ex and email. The notice explained the problem, risk, and requested the following actions be taken: Discontinue distribution of the product and dispose of all affected product, Distributors are directed to notify their customers. For further assistance please contact: 1-844-8BD- LIFE (1-844-823-5433) Say "Recall" when prompted; M-F 8am - 5pm CT
Details
- Recall number
- Z-1699-2023
- Initiated
- April 20, 2023
- Posted by FDA
- June 8, 2023
- Product code
- EZD – Catheter, Straight
- Quantity
- 7427880 devices
- Distribution
- Worldwide
- Recalling firm
- C.R. Bard Inc, Covington, GA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1698-2023 Class II | Magic3 Go Intermittent Urinary Catheters, Female Catalog Numbers: a) REF 51808, b) REF…Some catheters had a manufacturing defect where the clear resealing label separating from the product foil… | 04/20/2023Open, Classified |
| Z-1700-2023 Class II | Magic3 Go Intermittent Urinary Catheters, pediatric Catalog Numbers: a) REF 52810GSome catheters had a manufacturing defect where the clear resealing label separating from the product foil… | 04/20/2023Open, Classified |
| Z-1701-2023 Class II | HydroSil Go Intermittent Urinary Catheters, Catalog Numbers: a) REF 71808, b) REF 71810…Some catheters had a manufacturing defect where the clear resealing label separating from the product foil… | 04/20/2023Open, Classified |